E-mail
25946911@qq.com
-
Phone
18201857198
-
Address
No.1 Sanyou Road, Pujiang Town, Minhang District, Shanghai
Shanghai Zhilian Intelligent Technology Co., Ltd
25946911@qq.com
18201857198
No.1 Sanyou Road, Pujiang Town, Minhang District, Shanghai
ZH-JSQ100Dust particle counter sensorHaving higher measurement accuracy and particle size resolution, and stronger ability to resist the influence of humidity. Sensor particle counting sensor uses industrial grade laser and photosensitive components, combinedThe latest generationlight scatteringparticulate mattermonitorTechnology,Carefully tunedOptics andair circuitStructure,andNew generation high-frequencyweak signalcircuit processingandHigh precision particle recognition algorithm,moreaccurateofmeasurementvariousAn environment with cleanliness requirementsParticle concentration.
feature
lAccurate data: Single particle laser scattering principle, industrial grade laser source. High counting efficiency, good repeatability and consistency.
lParticle size range: Particle sizescope0.3~10μmWhile ensuring counting efficiency, a single sensor covers six channels.
lHigh reliability: aerodynamic correction,Optical and pneumatic self inspection, insensitive to vibration, ensuring system reliability and stability.
lStable flow: adopting active sampling method, using constant current intake fan, optionalPIDFlow control function, supporting stable, long-term, high negative pressure, and sampling; Optional air pump version, external high-performance air pump module.
lFast response: second level data updates.
lCommunication interface:RS485andUARTTTLTwo types of serial output; optional4GEthernet
lFully automatic measurement: can be remotely controlled, wirelessly transmitted, and real-time data uploaded to the Internet of Things platform.
lCircuit safety: equipped withESDProtection against overvoltage, overcurrent, and reverse connection ensures the stability and reliability of the circuit.
lEasy installation: The installation structure is simple and highly versatile.
lSafe and friendly: low voltage power supply, no electric shock; No radioactive sources, no light pollution.
lSampling flow rate: standard2.83L/min, air pump/There are two sampling methods available for the fan.
Scope of Application
medicineindustry electronics industry food processingindustry Precision machining clean room, etc.
working principle
ZH-JSQ100Dust particle counter sensorBy using the principle of single particle laser scattering, precise optical design and an air path designed based on fluid mechanics principles, particles in the sampled air are highly likely to pass through the beam in sequence, generating weak scattered light; Through precise optical signal receptionCollection deviceCollect scattered light and project it onto a high-sensitivity, high bandwidth photodetector. By identifying and analyzing the scattered pulses of each particle, the signal intensity corresponding to each particle is obtained; Obtain the particle size of each particle through a calibration program; After conversion and aerodynamic calibration, the number of particles is obtained.
technical indicators
serial number |
project |
parameter |
Remarks |
1 |
Measurement output |
0.3~10μmtogether6通道particle number |
0.3μmThe0.5μmThe1μmThe3μmThe5μmThe10μm |
2 |
range |
4, 000,000a /ft3 |
Complies with losses10%time |
3 |
input voltage |
DC12V [9~25VDC] |
Recommended power supplynot less than12V 2A |
4 |
rated power |
3.0W |
|
5 |
Sleep power |
0.24 W |
|
6 |
Working temperature range |
-20-60℃ |
|
7 |
humidity range |
0-99%RH |
|
8 |
Working atmospheric pressure |
86KPa~110KPa |
|
9 |
response time |
1S |
|
10 |
communication interface |
RS-485 + UARTTTL |
customizable |
11 |
communication protocol |
Modbus-RTU |
customizable |
12 |
Counting efficiency |
0.3μm/50%>0.45μm/100% (according toISO21501) |
@25℃,50%RH |
13 |
relative error |
±10% |
@25℃,50%RH |
14 |
Product dimensions |
168*67*82mm |
|
15 |
Hose size |
inner diameterφ5mmOuter diameter: φ8mm |
Excluding equal power sampling head |
16 |
lifespan |
3year |
40Below ℃ |