Welcome Customer !

Membership

Help

Changzhou Jinnan Instrument Manufacturing Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Products

Changzhou Jinnan Instrument Manufacturing Co., Ltd

  • E-mail

    jinnan@jinnan17.com

  • Phone

    13861129187

  • Address

    No. 5, Building 4, No. 39 Yangang East Road, Jincheng Town, Jintan District, Changzhou City, Jiangsu Province

Contact Now

Dual function water bath constant temperature oscillator

NegotiableUpdate on 12/20
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
SHA-B dual function water bath constant temperature oscillator, also known as water bath constant temperature shaker, is a biochemical instrument that combines a temperature controlled constant temperature water bath and oscillator. It is controlled by a microcomputer, with digital display and speed measurement display; When oscillating, there are small waves, but no splashing waves; Equipped with timing and constant speed settings, it runs smoothly and is easy and safe to operate.
Product Details
一、 SHA-BDual function water bath constant temperature oscillatorKey Features:
1. Microcomputer control, digital display and speed measurement display.
2. When oscillating, there are small waves, but no splashing waves.
3. Equipped with timing and constant speed settings, it runs smoothly and is easy and safe to operate.
4. The universal spring bottle rack is particularly suitable for the cultivation and preparation of biological samples for various comparative experiments.
5. The inner cavity is made of stainless steel, which has good corrosion resistance.
Second, SHA-BDual function water bath constant temperature oscillatorTechnical Specifications:
1. Power supply: 220V 50Hz
2. Heating power: 1800w
3. Timing range: 0-9999 minutes (166 hours and 39 minutes)
4. Oscillation frequency: 40-300 revolutions per minute, (can be set at a constant speed)
5. Oscillation amplitude: 20mm
6. Constant temperature range: room temperature -100 ℃
7. Oscillation method: reciprocating and swirling (dual function)
8. Temperature control accuracy:+0.5 ℃
9. Water tank size: 490 × 390 × 180 (mm)
10. Oscillatory plate size: 400 * 290 (mm) universal fixture
11. Special Reminder: Temperature/Speed/Time Display
3、 Notice before use
1. Please read this manual carefully before use.
2. The working surface of the machine should not be excessively smooth (such as tiles, etc.).
3. The electrical rated parameters of the socket provided by the user should not be less than the electrical rated parameters of the machine, and there should be good grounding measures.
4. The whole machine is strictly prohibited from being used in an environment with direct sunlight.
5. Set the speed according to the instructions. The speed has a call recovery function.
6. The machine may experience some displacement during high-speed oscillation, so it should be monitored by someone during use.
4、 Instructions for use:
1. Load the test bottle and maintain balance. If it is a dual function model, set the oscillation mode. (For detailed settings, please refer to the final instrument control panel diagram)
2. Connect the power and work according to the required time. Set the speed again
3. Turn on the power switch and set the constant temperature: To set the temperature, press the image.png key to set or view the temperature setting point. The character will start flashing, indicating that the instrument has entered the set state. Press image.png to increase, shift the image-png key or image-png to decrease, set the desired set temperature value, and then press the image-png key to return to standard mode. The heating indicator light will start to light up.
Please refer to the manual of the temperature control intelligent meter
Turn on the power switch, set the constant temperature: Set the temperature, press双功能水浴恒温振荡器The key can be used to set or view the temperature setting point. The character starts flashing, indicating that the instrument has entered the setting state. Press双功能水浴恒温振荡器Increase, shift双功能水浴恒温振荡器Healthy or双功能水浴恒温振荡器Reduce health, set the desired temperature value, and then press the set health button双功能水浴恒温振荡器Return to standard mode and the heating indicator light will turn on.
3. Press and hold for three seconds while working to enter the timer setting mode and set the timer time.
Down key (self-tuning key): Used as a decrease key in the setting state, long press for three seconds in the working state to enter self-tuning mode.
Up button: The high temperature alarm light is located at the bottom right corner. When it is on, it indicates that the high temperature has exceeded the limit, and when it is off, it indicates that the high temperature has not exceeded the limit.
Indicator light information:
AT light: The upper left is the self-tuning indicator light, flashing indicates self-tuning, off indicates normal working status.
TIMER light: The timing indicator light is located in the upper right corner. If it is off, it means the timer is turned off, and if it is on, it means the timer is turned on and working.
OUT light: The lower left is the heating indicator light. When it is on, it indicates heating, and when it is off, it indicates stopping heating.
ALM light: The high temperature alarm light is located in the lower right corner. When it is on, it indicates that the high temperature has exceeded the limit, and when it is off, it indicates that the high temperature has not exceeded the limit.
5、 Activate the oscillation device:
(1) Turn on the oscillation switch on the control panel, and the indicator light will turn on.
(2) Set the oscillation speed to the desired oscillation frequency. (Constant speed display)
(3) Cut off the power after work and set the temperature control to the low point.
(4) Clean the machine and keep it clean.
6、 Precautions
1. Instruments should be placed on a sturdy workbench, with a clean and tidy environment and good ventilation.
2. The power socket provided by the user should have good grounding measures.
3. It is strictly prohibited to move the machine during normal operation.
4. It is strictly prohibited for objects to collide with the machine.
5. Children are strictly prohibited from approaching the machine to prevent accidents.
6. Before replacing the fuse, make sure that the power supply has been cut off.
7. After use, please clean the machine without leaving any water droplets or dirt residue.