Doppler flowmeter
1、 Product principle:
In the field of acoustics, when there is relative motion between the sound source and the receiver (i.e. probe and reflector), the frequency of the echo will change, and this frequency change is called frequency shift, or Doppler effect. The frequency of sound waves increases when moving towards the observer and decreases when moving away from the observer. Doppler flowmeter calculates flow velocity by calculating frequency changes.
2、 Technical parameters of Doppler flowmeter:
Power supply:DC10-15V
Working current:160-180mA
Sound wave frequency:2MHZ
blind area:10centimeter
Flow rate range:0.021m/s~6.00m/s(Customizable)12.00m/s),
Accuracy: ±1%±1cm/s
Water level: Range:0.03m~10m(Customizable)100m)Accuracy:0.3%±0.5cm
Water temperature: Range:-10℃~60℃, accuracy: ±1℃
Instantaneous flow range:1L/s~99.99m³/s
Cumulative traffic range:0.1m³~999999m³
Working voltage:DC7.2V~24V
3、 Typical applications
open channel/Pipeline flow monitoring
Flow rate control in sewage treatment plants
ocean/hydrologic research
4、 Installation instructions:
Normal installation ensures that it is below the lowest water level20CMTo avoid exposure to water
Doppler flowmeterSensor installation diagram:

5 Wiring instructions:
Sensor red line |
power supplyDC12V+ |
Sensor black wire |
power supplyDC12V- |
Sensor Green Line |
RS485A |
Sensor yellow wire |
RS485B |
Notes:
1Sensors need to avoid mud and collisions with objects such as stones. Grids should be installed in areas with abundant water and grass to prevent entanglement.
2The installation direction of the sensor, the sensor emission surface needs to be installed in the upstream direction.
3The sensor has a blind spot for sound waves, so try to ensure that the sensor is underwater as much as possible20Below centimeters(Sensors that are partially exposed to water or obstructed by sludge may produce false data, with a very low probability. Typically, large numbers of data such as2rice/For seconds or more, it is recommended to use sensors. Secondary developers can block high flow velocity values by adjusting the water level, or in practical applications where the water flow velocity is not as high, they can directly block values greater than2Flow velocity above meters.
4The cable ventilation pipe is not difficult to break, which affects water level measurement.
5Try to do a good job in signal and power lightning protection.
6If the sensor cable needs to be extended, it needs to be added485Relay. The power supply should avoid interference from the frequency converter.