DK-0085A Differential Pressure Micro Moisture Analyzer
Main parameters:
Product features Temperature control range: 100 ℃~300 ℃
Temperature control fluctuation:+/-2.0 degrees Celsius
Vacuum degree: 100pa
Power supply: 220V 2A
Measurement range: 5-3000ppm (corresponding to a sample size of 3-0.5g, with a maximum sample size of around 10g)
After drying, trace amounts of moisture such as polyester chips and nylon 6 particles directly affect the spinning quality. At present, the main methods used for measuring trace moisture include differential pressure method, electrolysis method, Karl Fischer method, etc. Differential pressure method is fast, accurate, convenient, low-cost, and has a moderate instrument price, making it particularly suitable for factories and enterprises.
DK-0085A Differential Pressure Micro Moisture AnalyzerUsed for the determination of trace moisture in polyester chips and nylon 6 particles in the chemical fiber and plastic industries after drying treatment before melt spinning. Performance compliance: ISO6188-1980 "Plastics - Determination of moisture content in polyethylene terephthalate granules".
The instrument mainly consists of a differential pressure glass instrument, heater, automatic centering lifting device, temperature controller, vacuum gauge, vacuum pump, etc. It is equipped with sodium molybdate reagent, vacuum grease, silicone oil, vacuum hose, mercury, syringe, vacuum pump interface and all miscellaneous accessories and spare parts. Turn on the power, set the temperature controller to the desired temperature, raise the temperature and keep it constant; Pour the test sample into the test tube and insert it upwards into the test tube interface of the differential pressure glass instrument; Shake the right handle, raise the heater, and connect it with the test tube to perform the test according to the standard operating procedures.




Double helix stirring sodiumization machine, COSMO LS-R902 differential pressure leak detector from Japan