Calibration sheet for magnetic and eddy current coating thickness measuring instrument

NegotiableUpdate on 05/06
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Introduction to Calibration Plates for Magnetic and Eddy Current Coating Thickness Measuring Instruments $r $n Scope of Application: Suitable for calibrating various models of magnetic and eddy current coating thickness measuring instruments as standards. Applicable standard: JJG 818-2005 "Calibration Specification for Magnetic and Eddy Current Coating Thickness Measuring Instruments". Product Usage: The standard block is designed specifically for verifying magnetic and eddy current coating thickness measuring instruments, and is suitable for various models of magnetic and eddy current coating thickness measuring instruments.

Product Details

Calibration sheet for magnetic and eddy current coating thickness measuring instrument

Introduction

Scope of application:Calibration sheet for magnetic and eddy current coating thickness measuring instrumentA standard instrument suitable for verifying various models of magnetic and eddy current coating thickness measuring instruments.

Applicable standards:JJG 818-200Calibration Specification for Magnetic and Eddy Current Coating Thickness Measuring Instruments.

Product Usage: The standard block is designed specifically for verifying magnetic and eddy current coating thickness measuring instruments, and is suitable for various models of magnetic and eddy current coating thickness measuring instruments.

Technical Indicators:

Requirements for accuracy level, large allowable error of indication, and thickness sheet for calibrationμm

Project Name

Accuracy level

AA

A

B

C

D

Large allowable error of indication

±(0.3+0.5%H)

±(0.5+1%H)

±(1+3%H)

±(2+5%H)

±(3+10%H)

Calibration film

Uncertainty of thickness measurement results

h≤50

0.15

0.3

0.5

1.0

1.5

h>50

0.3%h

0.6%h

1%h

2%h

3%h

uniformity

h≤50

±0.15

±0.3

±0.5

±1.0

±1.5

h>50

±0.3%h

±0.6%h

±1%h

±2%h

±3%h

Note: 1H——The actual thickness value of the standard thickness sheet (or calibration plate coating);h——The actual thickness value of the calibration film.

2. The uncertainty of the thickness measurement result of the calibration sheet is the extended uncertainty,k=3.

Calibration sheet specifications (*)

11.9±0.3μm

23.3±0.3μm

48.6±0.3μm

99.0±0.3μm

243.7μm±0.6%

501.2μm±0.6%

996.5μm±0.6%

1985μm±0.6%

3805μm±0.6%

7546μm±0.6%

Note: One set comes with two substrates