COSMOSIL 5C18-PAQ chromatographic column

NegotiableUpdate on 05/15
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The COSMOSIL 5C18-PAQ chromatographic column can be analyzed with 100% water, making it suitable for analyzing organic acids and acrylamide that are highly hydrophilic and difficult to analyze on traditional C18 chromatographic columns.

Product Details

Product Description

COSMOSIL 5C18-PAQ chromatographic columnIt can be analyzed with 100% water, making it suitable for analyzing organic acids and acrylamide that are highly hydrophilic and difficult to analyze on traditional C18 chromatography columns.

  • 100% water can be used as the mobile phase

  • Suitable for hydrophilic compounds

COSMOSIL 5C18-PAQ chromatographic column

Stability under 100% water flow phase

The following figure shows the change in retention time of thymine under 100% aqueous phase (20 mmol/l phosphate buffer pH 7). After the sample was analyzed 4 times, stop transporting the mobile phase for 1 hour. When analyzed again under the same conditions, the retention time of the traditional C18 column changed before and after stopping the delivery, while the retention time of COSMOSIL 5C18-PAQ remained stable.


Application Cases

COSMOSIL 5C18-PAQ应用案例


Usage information

chromatographic column 5μm C18-PAQ
Inner diameter (mm) 1.0 2.0 3.0 4.6 6.0 10 20 28
Column length (mm) ALL ALL ALL ALL ALL ALL ALL ALL
Factory solvent Acetonitrile/water=50/50 Methanol/water=66/34 Methanol/water=80/20
flushing method
  1. Method for removing buffer, salt, and acid from the chromatographic column: Rinse with a mobile phase free of buffer, salt, and acid for 10-15 minutes.
    Example: When the mobile phase is methanol/phosphate buffer (20mmol/L)=50/50, rinse with methanol/water=50/50

  2. The method of flushing the attached substances in the chromatographic column and the solution when the baseline is unstable

  • When the sample is not protein, rinse with methanol or *.

  • When the sample is protein, use 0.1%C2HF3O2Mix 50-70% acetonitrile/waterwash

保存方法
  1. Short term (several days) storage:
    Rinse with a mobile phase without buffer, salt, or acid for 10-15 minutes. preservation.
    Example: When the mobile phase is methanol/phosphate buffer (20mmol/L)=50/50, rinse with methanol/water=50/50

  2. Long term (over 1 month) storage:
    After use, clean the chromatography column with a mobile phase free of acid and salt, then replace the mobile phase with acetonitrile/water=70/30 or methanol/water=70/30, and store.

pH range pH2~pH7.5
pressure resistance 20MPa 15MPa
temperature range Highly resistant to 60 degrees, it is recommended to analyze at 20-50 degrees.
Applicable solvents

It can be used except for alkaline solutions and strong acid solutions (pH below 1.5).(Strong alkaline solution will dissolve silica gel, strong acid solution will cause the stationary phase to detach)
Points to note:
Excessive solvent viscosity can cause pressure rise, please control the pressure below 20Mpa.After using acidic mobile phase or high freezing point solvents, the chromatography column must be cleaned.

Precautions
  • In general, a buffer concentration of 0.005~0.02 mol/L is sufficient.

  • The mobile phase must be filtered through a membrane below 0.5 μ m before use.

  • When analyzing proteins in reverse phase mode, please use a large pore size chromatography column (pore diameter 300 Å)COSMOSIL Protein - R(粒径5μm)。



Shenzhen Dongfang Lantai Technology is a high-tech company specializing in the import of laboratory analytical instruments and equipment such as gas chromatography, liquid chromatography, electronic balance, electrochemical instruments, general instruments and equipment, chromatography columns, reagents, standards, etc. At the beginning of its establishment, the company insisted on focusing on the analysis industry and was committed to introducing domestic and foreign analytical instruments to a wide range of domestic users. Currently, the company's users have spread to scientific research institutions, colleges and universities, hospitals, chemical industry, food, medicine, agriculture, mining, power, metallurgy and other industries and units.