Phone
18015580277
-
Address
No. 55 Dongfu Road, Suzhou Industrial Park
Suzhou Yucheng Instrument Co., Ltd
18015580277
No. 55 Dongfu Road, Suzhou Industrial Park
LY-09Introduction to WBGT Index Meter
OneBlack ball wet bulb index meterPurpose and principle
Black ball wet bulb index meterIt is used to evaluate the meteorological conditions of high-temperature workshops, taking into account four factors: air temperature, wind speed, air humidity, and radiant heat.WBGTIt is composed of three temperature components: black sphere, natural wet sphere, and dry sphere. The temperature reading of the black sphere is affected by air temperature, radiation heat, and wind speed, while the natural wet sphere temperature is affected by air temperature, wind speed, radiation heat, and air humidity. After long-term experimental research, it is believed that when these three readings are comprehensively calculated using the following method, the results obtained can accurately reflect the meteorological conditions of the workplace.
WBGTThe calculation method of the index is as follows:
When there is no solar radiation heat indoors and outdoors:
WBGT=0.7tnw+0.3tg
When there is solar radiation heat outdoors
WBGT=0.7tnw+0.2tg+0.1ta
When calculating various parameters, the average value should also be taken into account as it varies with time and air.
IIWet bulb black bulb temperature WBGT index meterStructure of the instrument
Wet bulb black bulb temperature (WBGT) index meterIt consists of two parts: sensor probes and indicator instruments.
1Measuring head
Three sensors measuring natural wet bulb temperature, black bulb temperature, and dry bulb temperature.
⑴ Natural wet bulb temperature probe
Signal conditioning |
|
A/D converter |
|
To all levels |
Measurement range: 5-40 ℃
Accuracy: ± 0.5 ℃
⑵Black ball temperature sensor
Measurement range: 20~120 ℃
Accuracy: 20-50 ℃± 0.5 ℃; 50~120℃ ±1℃
⑶Air temperature sensor
Measurement range:10~60℃
Accuracy:±0.5℃
The digital temperature sensitive components used in this instrument ensure the accuracy and stability of the testing.
2.Indicator instrument
(1)Display and Reading
Turn on the power switch, and the instrument will directly display the indoor measurementWBGTOutdoorWBGTThe results include wet bulb temperature, black bulb temperature, dry bulb temperature, and time.
(2)Data playback
The measured data can be replayed. When replaying, turn off the device first, press and hold the replay button to turn it on and enter the replay program. The measured data is stored by stacking, and the delayed data is played back first. Then, press the middle button (next number) once to display the previous set of data.32The group data is displayed alternately.