Biological safety cabinet BSC-1004IIA2/BSC-1304IIA2, color high-definition LCD human-machine dialogue interface, real-time display of monitoring fan, filter and other operating status and providing alarm function, filter life display, etc. Efficient and energy-saving DC brushless constant airflow fan; Mobile operating table, equipped with stainless steel lifting handle and support frame, making cleaning and maintenance more convenient.
Biosafety Cabinet BSC-1004IIA2/BSC-1304IIA2
Product Description:
biological safety cabinetIt is a biosafety isolation equipment used in microbiology, biomedical, biosafety laboratories, and other laboratories. It adopts advanced air purification technology and negative pressure box design, which realizes the protection of the environment, personnel and samples, and can prevent the diffusion of harmful suspended particles and aerosols; Providing safety protection for operators, samples, and cross infection between samples, as well as the environment, is a basic safety protection equipment in the first level biological safety barrier of the laboratory. Adopting a bracket structure, the bracket and the box can be separated for easy handling and positioning.
Product features:
1. Friendly human-machine dialogue interface
◆ Color high-definition LCD human-machine dialogue interface, real-time display of monitoring fan, filter and other operating status, providing alarm function, filter life display, etc.
2Efficient and energy-saving DC brushless constant airflow fan
◆ Energy saving of over 40% year-on-year, and the use of constant air flow fans and intelligent air volume compensation systems ensure that the air volume changes by less than 10% when the resistance of the high-efficiency filter increases by 50%, making it safer to use.
3Humanized structural design
The front window glass can be fully cleaned both inside and outside, completely solving the problem of internal uncleanliness and leaving no hygiene blind spots.
Mobile operating table, equipped with stainless steel lifting handle and support frame, making cleaning and maintenance more convenient.
Anti splash safety sockets are installed on both sides of the inner wall of the work area, and two water and gas valve connection channels are installed on each side for more flexible use.
◆ Adjustment feet without exposed threads.
4Power off memory function
When the experimental process needs to be paused, there is no need to shut down. Closing the front window will automatically enter a low-speed energy-saving operation state, maintaining the cleanliness of the operating area. Once the front window is opened, it immediately enters normal operation mode, effectively improving operational efficiency. If there is an unexpected power outage, it can automatically return to the operating state before the power outage after restoring the power supply, fully protecting personnel safety.
5. High quality ULPA filters
Adopting the high-efficiency, low pressure drop, high-strength, and low boron content air filter element from HV Company in the United States, the service life of the filter is extended while reducing pressure drop.
Both the air supply and exhaust filters are equipped with "leak proof" technology to ensure air cleanliness at ISO level 4 (level 10).
The ULPA filter has a retention efficiency of 99.9995% for solid particles with a diameter greater than 0.12 microns.
Technical Parameter
Model/Parameters |
BSC-1004IIA2 |
BSC-1304IIA2 |
BSC-1604IIA2 |
BSC-1804IIA2 |
Filtration efficiency |
≥99.9995%, @0.12μm |
Air supply and exhaust filters |
ULPA filter |
| air cleanliness |
ISO 4 |
| Product execution standards |
YY0569(GB 4793.1、GB/T 18268.1) |
Decreased wind speed (m/s) |
0.35 |
0.33 |
Inflow wind speed (m/s) |
0.55 |
0.53 |
Noise (dB (A)) |
58~65 dB(A) |
| Vibration (μ m) |
The center position of the tabletop is ≤ 5 |
power supply |
AC220V,1φ,50Hz |
| Opening height of sliding door (mm) |
200 |
| Cross infection prevention |
A. The total bacterial count of the impact sampler is ≤ 10CFU per time
B. The total bacterial count of the slit sampler is ≤ 5CFU per time
|
| Personnel protection |
Total colony count ≤ 5CFU./time |
| Protection of the tested product |
Total colony count ≤ 2CFU./time |
|
ZHigh power (including backup socket) (kVA) |
1.65 |
1.75 |
1.85 |
1.85 |
| Rated power (excluding spare socket) (kVA) |
0.33 |
0.43 |
0.53 |
0.53 |
Net weight of product (Kg) |
260 |
300 |
360 |
400 |
Internal dimensions (width X depth X height) (mm) |
1004*630*630 |
1304*630*630 |
1604*630*630 |
1804*630*630 |
External dimensions (width X depth X height) (mm) |
1200*795*2050 |
1500*795*2050 |
1800*795*2050 |
2000*795*2050 |
Lighting specifications (W) and quantity |
LED 24.5*① |
LED 31*① |
LED 36*① |
LED 36*① |
UV lamp specifications (W) and quantity |
18*① |
30*① |
36*① |
36*① |
Illuminance (Lx) |
≥900 |
| Outer box structure |
High quality cold-rolled steel plate electrostatic powder coating |
| Workspace Structure |
High quality 304 stainless steel one-time molding |
Exhaust direction |
eject |
| |