Benzene series automatic samplerMain structure:
The multifunctional gas sampler mainly consists of a multi way valve, a ten way valve, an EFC electronic mass flow meter, an inerting pipeline system, a time and valve control unit, a GC trigger and sampling pump control unit, a touch screen, a connection device, etc.
Working Principle:
The multifunctional gas sampler controls the flow rate and volume through a sampling pump and EFC electronic mass flow meter, takes timed samples, and then uses GC carrier gas to bring the sample into the GC chromatographic column and trigger GC for analysis by switching the ten way valve.
Benzene series automatic samplerMain performance parameters:
1. Input power supply voltage: single-phase AC220V ± 10%, 50HZ
2. Power: 0.5KW (AC)
3. Standard gas sampling flow rate: 0~100ml/min
4. Sample sampling flow rate: 0~500ml/min
5. Dilution ratio: ≤ 500 times
6. Injection volume: 1ml
7. Sampling time range: 0~999 seconds
8. Temperature range for use: room temperature -50 ℃
9. Injection method: Quantitative cup
10. Number of quantitative rings: 1ml * 2