Welcome Customer !

Membership

Help

Xiamen Xinrui Instrument Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Products

Xiamen Xinrui Instrument Co., Ltd

  • E-mail

  • Phone

  • Address

    Room 1015, Building A, MediaTek Electronics Market, No. 800 Yuanshan South Road, Xiamen City

Contact Now

BT3554-51 daily HIOKI battery internal resistance detection

NegotiableUpdate on 05/06
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Product Overview: $r $n $r $n ● Even if the battery is installed on the device, it can be determined without power outage. $r $n ● The fastest time from measurement to record saving is only 2 seconds, which is 60% shorter than the old model 3554. $r $n ● Test internal impedance/voltage, instantly determine aging degree PASS/Warning/FAIL * 1 $r $n ● Equipped with noise reduction technology to improve noise resistance $r $n ● Simplify testing steps based on host interface and voice prompts * 2 $r $n ● Test data can be associated and saved with information on the measurement site, thereby reducing management time. $r $n ● All kinds of measurement data can be used with GE

Product Details

Product Overview:

Even if the battery is installed on the device, it can be determined without power outage
The fastest time from measurement to record saving is only 2 seconds, which is 60% shorter than the old model 3554.
Test internal impedance/voltage and instantly determine the degree of aging PASS/Warning/FAIL * 1
Equipped with noise reduction technology to improve noise resistance
Simplify the testing steps based on the host interface and voice prompts
Test data can be associated and saved with information from the measurement site, thereby reducing management time.
All types of measurement data can be uniformly managed using GENNECT Cross software
The measured values are transmitted directly to Excel via wireless means ® In the form (requires optional wireless adapter Z3210)
Equipped with protectors to improve portability and durability for on-site use.

The threshold for qualified judgment varies depending on the manufacturer, type, capacity, etc. of the battery. The internal impedance/terminal voltage of new or qualified batteries needs to be measured in advance. Compared with sealed lead-acid batteries, open type (liquid type) lead-acid batteries or alkaline batteries may sometimes be difficult to determine aging due to their less variation in internal resistance.
※2 Bluetooth ® After connection, the terminal will have a voice prompt.
Using tablet or smartphone applications can achieve data recycling. (Requires optional wireless adapter Z3210)

HIOKI/Japanese BT3554-50/BT3554-51Battery TesterBasic parameters (accuracy guarantee time of 1 year)

model

BT3554-50 (host model, not sold separately)

BT3554-51 (including host BT3554-50, needle test line 9465-10)

BT3554-52 (including host BT3554-50, needle test line L2020)

Resistance measurement range 3 m Ω (maximum display 3.100 m Ω, resolution 1 µ Ω) to 3 Ω range (maximum display 3.100 Ω, resolution 1 m Ω), 4-speed switching
Measurement accuracy: ± 0.8% rdg ± 6 dgt (only 3 m Ω range ± 1.0% rdg ± 8 dgt)
Measure current frequency: 1 kHz ± 30 Hz, when the noise frequency avoidance function is turned on, 1 kHz ± 80 Hz
Measurement current: 160 mA (3 m Ω/30 m Ω range), 16 mA (300 m Ω range), 1.6 mA (3 Ω range)
Open terminal voltage: 5 V Max
Voltage measurement range ± 6 V (maximum display ± 6.000 V, resolution: 1 mV)~± 60 V (maximum display ± 60.00 V, resolution: 10 mV), 2-level switching, measurement accuracy: ± 0.08% rdg ± 6 dgt
Temperature measurement accuracy Measurement range -10 ° C to 60 ° C, maximum display 60.0 ° C, resolution 0.1 ° C, measurement accuracy ※ ± 1.0 ° C
When using the clamp type temperature sensor 9460
Accuracy of BT3554-50 monomer: When simulating input, a clamp type test line 9460 with an attached ± 0.5 ° C temperature sensor is required
When using temperature probe 9451, please add ± 0.5 ° C (cable length 1.5m)
When using temperature probe 9451s, please add ± 0.5 ° C (cable length 0.1m)
Accuracy of BT3554-50 monomer: ± 0.5 ° C for analog input
Maximum allowable input voltage DC 60 V Max. Unable to input AC
measure time 100 ms
response time About 1.6 seconds
comparator

Comparison and notification of set thresholds and measured values
Determination notification method: The results in the table below are displayed in segments and beep sound
Voltage value (high): Impedance value (low)=PASS, Impedance value (medium)=Warning, Impedance value (high)=FAIL
Voltage value (low): Impedance value (low)=Warning, Impedance value (medium)=Warning, Impedance value (high)=FAIL
When the judgment result is "Warning" or "FAIL", the buzzer will sound and the screen will turn red
backlight
Voltage determination method: ABS (absolute value determination), POL (± polarity determination)
Save settings: 200

Storage function Operation content: Save, read, and delete measurement data
Saving and deleting preset information
Data volume: 6000
Memory composition: Each unit stores 500 pieces of data (12 units)
Save content: Measurement data and preset information are saved together
① Measurement data
By operating the host, you can save, read, and delete
-1. Date and time
-2. Impedance value, voltage value, temperature
-3. comparator threshold and judgment result
2. Preset information
Save, read, and delete from the corresponding application software (GENNECT Cross/GENNECT ONE)
-1. Data Number: Numbers 1 to 100
※ Cannot save the same number
For one data number, it will be saved in sequence three four
-2. Location information: When using UPS, it can be annotated arbitrarily
-3. Equipment information: UPS management number, etc., can be annotated arbitrarily
-4. Battery number: 1-500 (start number, end number)
Measurement Record Guidelines Operation content
Display and voice notify the next measured battery number through the host interface
※ Voice is output through smartphones/tablets when used through Z3210 and application software (GENNECT Cross).
Preparation in advance: Log in to the corresponding application software (GENNECT Cross/GENNECT ONE)
Host transmits preset information
communication interface USB
Bluetooth ®️ Wireless communication (when equipped with Z3210)
Other functions Temperature measurement (-10.0 ℃~60.0 ℃), zero adjustment, hold function, automatic hold function, automatic storage function, automatic power-saving mode, clock
power supply 5 alkaline battery (LR6) x 8
Continuous usage time: approximately 8.3 hours (without Z3210)
About 8.2 hours (when equipped with Z3210 and wireless communication)
Volume and weight 199 W × 132 H × 60.6 D mm (when installing the protective shell)
960 g (including battery and protective case)
Attachment Carrying case C1014 × 1, protective shell Z5041 × 1, fuse set Z5050 × 1, zero adjustment board × 1, hanging rope × 1, USB cable × 1, GENNECT ONE software CD × 1, 5 # alkaline battery (LR6) × 8, user manual × 1
No attachment probe Needle shaped testing line 9465-10 × 1 Needle shaped testing line L2020 × 1