Welcome Customer !

Membership

Help

Beijing Kangtai Electronics Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Products

Beijing Kangtai Electronics Co., Ltd

  • E-mail

  • Phone

  • Address

    17 Qinghua East Road, Haidian District, Beijing

Contact Now

BMA-400 Four Channel Low Noise Bioamp

NegotiableUpdate on 05/10
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

BMA-400 four channel low-noise biological amplifier, a four channel biological signal amplifier that can be used for electroencephalography, electrocardiography, electromyography, and other extracellular applications. The BMA-400 has a low noise, wide gain range, and sharp cut-off bandpass filter, which can be used as a stable DC amplifier and provides an adjustable DC input bias control. Logic input control is achieved by recording the stimulation of electrodes. The stimulation/recording input and four channel output are connected to the multi pin connector on the rear panel for easy cable connection. Four standard input cables are provided with the device, along with a series of electrode leads (needles, buckles, etc.).

Product Details

Product Description: How to Translate

BMA-400 Four Channel Low Noise BioampIt can be used for electroencephalography, electrocardiography, electromyography, and other extracellular applications. The BMA-400 has a low noise, wide gain range, and sharp cut-off bandpass filter, which can be used as a stable DC amplifier and provides an adjustable DC input bias control.

Logic input control is achieved by recording the stimulation of electrodes. The stimulation/recording input and four channel output are connected to the multi pin connector on the rear panel for easy cable connection. Four standard input cables are provided with the device, along with a series of electrode leads (needles, buckles, etc.). Unprocessed input plugs can be used for customized electrode configurations.

BMA-400 Four Channel Low Noise BioampThe applications include multi-channel EEG, EMG, and other extracellular biological potential measurements. Head stage can be used for high impedance electrodes or customized recording, including ISO-Z isolated head stage.

Product advantages:

·4-channel AC/DC amplifier, suitable for extracellular data recording

·Low noise, wide gain range

·Multi channel EEG, EMG, ECG recording

·True DC response with bias

·Stimulus transmitted to electrodes

·Optional isolated head stage (ISO-Z) and high impedance head stage (HS-1)


Product specifications:

·Input type: differential

·Input range corresponding to full-scale output:+100mV -0.5V
·Input impedance:>1000 Megohm
·Broadband noise, RTI:<7 μ V P-P,
·Common mode rejection:>95dB
·Gain range: 10 X-50000X, 12 segments
·Low pass filtering: 100Hz-50kHz, 6-segment
·High pass filtering: DC-100Hz, 6-segment

·Filter type: Butterworth, flattest response, -12dB/octave
·Trigger/Record Input: TTL
·Input connector (rear panel): Amphenol 703-91T-3478-009 7-pin
·Output connector (rear panel): BNC socket
·Data port connector (rear panel): 25 pin D-SUB female
·Power supply requirements: 115/230VAC, 10VA
· 尺寸: 12W x 4H x 9D inches, 30 x 10 x 23 cm

Note: Not suitable for human life support applications.