Automatic industrial analyzers are suitable for industries such as power, coal, metallurgy, petrochemicals, environmental protection, cement, papermaking, geological exploration, and scientific research institutions to determine the moisture, ash content, and volatile matter of coal, coke, or other organic matter, and calculate their fixed carbon and calorific value. They can also analyze fly ash, ash residue, coke, and carbon content. Adopting standard: GB/T212-2008 "Method for determination of calorific value of coal"
Automatic industrial analyzerScope of application:
Suitable for industries such as electricity, coal, metallurgy, petrochemicals, cement, papermaking, geological exploration, and research institutions to determine the moisture, ash content, and volatile matter of coal, coke, or other organic matter, and calculate their fixed carbon and calorific value. It can also analyze fly ash, ash residue, coke, and carbon content.
Automatic industrial analyzerAdopting standard: GB/T212-2008 "Industrial Analysis Methods for Coal".
Automatic industrial analyzerThe main characteristics are:
1. Standardized process and accurate results: The testing process, conditions, and results for moisture, ash content, and volatile matter all meet the standard requirements and can meet the testing requirements for all types of coal without calibration, and can be used for arbitration analysis.
2. Fast testing speed: Adopting a multi furnace structure with a constant temperature drying device, the three indicators of moisture, ash content, and volatile matter can be measured in any combination or separately. At the same time, testing the three indicators of 12 samples can be completed within 150 minutes.
3. High degree of automation: RS485 on-site industrial control bus is used for communication, high-precision electronic balance is integrated into the instrument, combined with a set of pneumatic control automatic weighing mechanism, using thermogravimetric analysis method to automatically weigh, send, process data, calculate results, print and store reports, etc. The fixed carbon content, hydrogen value and calorific value can be automatically calculated by measuring moisture, ash content and volatile matter.
4. Easy to operate: During the experiment, there is no need to remove or place crucible covers, send samples, or take samples, avoiding the risk of high temperature radiation and burns.
5. High cost-effectiveness: Using a single testing host to achieve precise measurement of three industrial analysis indicators according to standards, with high cost-effectiveness and complete functions.
6. Dynamic addition of samples: During the testing process, samples can be temporarily added to adapt to situations where the user's sample is uncertain.
Automatic industrial analyzerTechnical parameters:
1. Sample weight: 0.5-1.5g
2. High operating temperature: 1000 ℃
3. Temperature control accuracy: ± 1 ℃
4. Sample quantity: 1 to 12 samples of water, ash, and volatile are made simultaneously
Otherwise, 1-24
5. Test temperature: 105 ℃ (moisture)
815 ℃ (ash content)
900 ℃ (volatile matter)
6. Testing time: 12 samples, three indicators testing time ≤ 150 minutes
7. Precision: Complies with GB/T212-2001 standard
8. Accuracy: within the allowable deviation range of the standard sample
9. Power: ≤ 4kW
10. Power supply: (220 ± 22) VAC, (50 ± 1) Hz
11. Dimensions: 620X530X760mm
12. Weight: 65Kg