OilLab 8611 automatic aniline point analyzer is used to detect the aniline point and mixed aniline point indicators of petroleum products and hydrocarbon solvents. Test method E describes the use of automated instruments to measure tests including methods A and B.
OilLab 8611 Automatic Aniline Point Tester meets the standards of D611-E, IP2-E related ASTM D611-A, B, C, D, IP 2-A, B, C, and D. It is used to detect the aniline point and mixed aniline point indicators of petroleum products and hydrocarbon solvents. Test method E describes the use of automated instruments to measure tests including methods A and B.
1、 Technical features:
1. OilLab 8611 automatic aniline point analyzer can automatically measure samples with ASTM chromaticity values less than 8;
2. The sample tube system can automatically lift and lower;
3. Electronic heating control system, achieving precise temperature control through PID algorithm, with built-in semiconductor liquid heat exchange system, capable of temperature control from -10 ℃ to 160 ℃;
4. Heating/cooling metal bath, safer;
5. Detachable glass sample tank, easier to clean;
6. The operator does not need to come into contact with aniline, which is safer. There is a small hole on the mobile head, and the sample can be injected into the sample pool through a 10 ml syringe; img1
7. Built in 8-inch touchscreen computer, equipped with dedicated aniline point management software, built-in standard heating and cooling program files and user-defined files;
8. Temperature and fiber optic detection probes are directly inserted into the sample for detection, rather than outside the sample pool;
9. Brass material mixer;
10. Three dimensional structure with anti-corrosion coating;
11. Dual detection system, capable of detecting dark and clear samples;
12. Built in touch screen computer:
(1)TFT/LCD, 8-inch
(2) 40GB hard drive
(3) Resolution: 1024x768, 16M
(4) 2 USB ports, can be used to connect external printers or other external devices;
13. Dedicated management software, bath temperature controlled by PT100 sensor, can display ℃ or ℉, with over temperature safety alarm;
14. The operating software can monitor and record the following information in real-time:
(1) Curve changes during the testing process
(2) Error testing prompt
(3) Output files can be provided in formats such as. xls/. pdf/. png
(4) Up to 100 calibration points can be reached;
15. If a lower temperature, such as -40 ℃, is required, an external cold bath can be selected.
2、 Technical parameters:
1. Aniline point measurement range: -15 ℃ to+160 ℃.
2. Temperature resolution: 0.01 ℃.
3. Temperature accuracy: ± 0.1 ℃.
4. Temperature repeatability/reproducibility: meets or exceeds the requirements of standard methods.
5. Sample type to be tested: transparent, turbid, or black sample.
6. Temperature sensor: PT100.
7. Aniline point detector: photodetector, dual detection unit.
8. Sample measuring cup: U-shaped tube, vacuum jacket, easy to clean.
9. Heating unit: electric heating jacket, PID control.
10. Refrigeration unit: Built in semiconductor refrigeration and liquid heat exchange system.
11. Test head: capable of automatic lifting.
12. Stirring element: gentle stirring, driven by an automatic electric motor.
13. Built in computer and LCD touch screen.
14. Built in powerful operating software.
3、 Other:
1. Power supply: 220V ± 15%, 50/60Hz
2. Environmental requirements: High temperature of 35 ℃; *High humidity of 80%
3. Size: 53 × 31 × 75 cm
4. Weight: 30kg