- Phone
-
Address
No. 169 Xuesong Road, Zhengzhou National High tech Industrial Development Zone
Hanwei Technology Group Co., Ltd
No. 169 Xuesong Road, Zhengzhou National High tech Industrial Development Zone
AT8000 is a newly developed portable exhaled gas alcohol content detector by our company, which is currently one of the most advanced internationally. The core component of this instrument adopts a new high-tech electrochemical alcohol sensor, which can accurately detect the alcohol content in exhaled gas and has strong anti-interference ability; The AT8000's streamlined keyboard, menu operation, and human-computer interaction interface make operations more concise and convenient; Color LCD screen display makes the interface more beautiful and clear; Built in wireless module, realizing wireless printing function; The disposable anti backflow nozzle ensures the health and safety of the test subjects.
This product has been shortlisted for the Traffic Management Equipment Procurement Catalogue of the Police Equipment Procurement Center of the Ministry of Public Security. The product design, manufacturing, and inspection shall comply with the following national standards:
GB/T 21254-2007 "Alcohol content detector for exhaled gas"
GB 19522-2004 "Thresholds and Testing of Blood and Breath Alcohol Content for Vehicle Drivers"
This product is suitable for special departments such as police and security to detect drunk driving or alcohol testing before work, effectively avoiding accidents and ensuring the safety of people's lives and property.
| Product Model | AT8800 |
| sensor type | Electrochemical alcohol sensor |
| measurement range | 0~2.000mg/L |
| Critical value for drinking and driving | 0.090 mg/L |
| Critical value for drunk driving | 0.363 mg/L |
| battery parameters | Two AA nickel hydrogen batteries with a rated capacity of 2.1AH |
| Recommended charging time | 12 hours |
| Continuous working hours | Normal detection status is greater than 48 hours |
| Operating Temperature | -10℃~50℃ |
| Operating Humidity | <95% RH (no condensation) |
| Shell protection level | IP55 |
| host size | 122mm×67mm×31mm |
| host weight | About 120g |
| Alcohol gas concentration range C (mg/L) | error |
| C<0.400 | ±0.020mg/L |
| 0.400≤C<1.000 | ±8% |
| C≥1.000 | ±30% |