AND EK-15KL Industrial Precision Electronic BalanceDisplay accuracy is0.1g, Range15kgRepetitive accuracy0.3gUltra wide and wide range, high protection design, stainless steel scale pan corrosion-resistant, easy to clean. Large font with backlightLCDDisplay, the reading is clear and visible.

Function:
·EC-15KLMeasurements with strict requirements for high resolution and large range are particularly suitable
·EC-30KLWhen the gross weight of the measured object exceeds15kgBut the net weight measurement reading needs to reach0.1gEspecially suitable for time
·Oversized scale pan(300×210mm)Suitable for weighing large objects or quantities
·Corrosion resistant stainless steel(SUS304)Scale pan
·Easy to read with backlightLCDDisplay screen(22mmHigh characters)
·Multiple weighing units:g,kg,oz,oztCounting mode(pcs), percentage mode(%)
·Maintain functionality, accumulate functionality(M+)Automatic power shutdown function
·ByRS-232CThe interface can output data to external devices such as computers and printers
·Optional AC adapter or rechargeable battery pack for power supply
Specification parameters:
model |
EC-15KL |
EC-30KL |
|
range |
15 kg |
3 kg / 30 kg |
|
minimum reading |
0.1g |
0.1 g / 1 g |
|
Counting Mode |
Minimum reading unit |
0.1g |
|
Sample Quantity |
5, 10, 25, 50 or 100 |
||
Percentage mode |
minimum100%display unit |
0.1% |
|
minimum100%weight |
10 g |
||
Weighing disk size |
300 × 210 mm |
||
useAND EK-15KL Industrial Precision Electronic BalancePrevious precautions:
A suitable location for placing scales, for20Temperature and humidity50%In the environment;
Do not place the balance in direct sunlight or near air conditioning;
Do not place the balance in dusty areas;
Do not place the balance in places with strong magnetic fields;
Please place the balance in the place where it is least likely to shake;
Do not place the balance on a storage platform that is prone to shaking;
Adjust the bubble to the center position within the red circle of the level using the horizontal adjustment footrest;
Please confirm whether the transformer matches the voltage of the usage location before use;
Before use, please plug the balance into the electrical appliance and preheat it30minute
Before use, the balance needs to be calibrated after being moved.