Overview

Introduction: Image model 8103A (aluminum shell) 8103B Measurement range 0-100 g/L; 0-20mg/L temperature range 0-60 ℃ polarization time 60min polarization voltage 0.7V accuracy 0.01mg/L drift 2%/month detection lower limit 0.1mg/L breathable film thickness 50m 100m shell material 316L stainless steel/PVC electrode current (air) 50-80nA I 0.1 ppm (note: maximum) ..

Product Details

image 8103A型法极谱式溶解氧电极 8103B型法极谱式溶解氧电极
model 8103A (aluminum shell) 8103B
measurement range 0-100 g/L; 0-20mg/L
temperature range 0-60℃
Polarization time 60min
polarization voltage 0.7V
precision < ± 0.01mg/L
drift <2%/ mounth
Detection lower limit <0.1mg/L
Thickness of breathable film 50µm 100µm
Shell material 316L stainless steel/PVC
Electrode current (air) 50-80nA I 0.1 ppm (note: maximum current 3.5 µ A)
thermistor Pt1OO/Pt1000/22K/2.252K
calibration 90 days
cable Standard 5m (customizable)
Scope of application Urban sewage treatment, industrial sewage treatment, aquaculture, and environmental monitoring