8101 series online dissolved oxygen electrode (ppb level)

NegotiableUpdate on 11/09
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Introduction: Principle and Characteristics of 8101 Series Online Dissolved Oxygen Electrode DOG-208F Type Dissolved Oxygen Electrode Application Polarographic Principle Platinum (Pt) as cathode, Ag/AgCI as anode electrolyte 0.1M chloride charge (KCI), imported silicone rubber permeable membrane as breathable membrane, with the properties of collision resistance, corrosion resistance, high temperature resistance, and no deformation of silicone rubber and steel mesh. 8101 Series Online Dissolved Oxygen Electrode Picture 8101

Product Details

Principle and Characteristics of 8101 Series Online Dissolved Oxygen Electrode

Application of Polarographic Principle for DOG-208F Dissolved Oxygen Electrode

Using platinum (Pt) as the cathode and Ag/AgCI as the anode

The electrolyte is 0.1M KCI chloride

Using imported silicone rubber permeable membrane as a breathable membrane

Featuring silicone rubber and steel mesh

Resistance to collision, corrosion, high temperature, and deformation

8101 series online dissolved oxygen electrode image

8101系列在线溶解氧电极(ppb级)
8101系列在线溶解氧电极(ppb级)剖面图

8101系列在线溶解氧电极(ppb级)流通池尺寸

8101 series online dissolved oxygen electrode technical parameters

Model: 8101
Measurement range: 0-100 g/L; 0-20mg/L
Electrode material: 316L stainless steel
Thermistor: 2.252K、22K、Pt100/1000
Sensor design lifespan: Over 3 years
Cable length: Standard 5m (double shielded)
Detection limit: 0.1 μ g/L ( ppb ) ( 20℃)
Detection limit: 20 mg/L(ppm)
Response time: ≤3min ( 90%, 20℃)
Polarization time: >8h
Polarization voltage: 0.7V
Minimum flow rate: 5cm/s; 5 15L/h
Drift: <3%/month
Measurement error: <1 ppb
Current in the air: 50-80nA note: maximum current 20-25uA
Zero oxygen: <5ppb(60min)
Calibration interval time: >60 days
Medium temperature: 0-60℃

Application scope of 8101 series online dissolved oxygen electrode

Applied to the measurement of trace oxygen content in thermal power plants, desalinated water in power plants, boiler feedwater, etc