E-mail
jcybok@163.com
-
Phone
13262809751
-
Address
Building 6, Lane 759, Xinye Road, Qingpu Industrial Park, Shanghai
Jicheng Instrument (Shanghai) Co., Ltd
jcybok@163.com
13262809751
Building 6, Lane 759, Xinye Road, Qingpu Industrial Park, Shanghai
The 6302A PT100 platinum resistance thermometer is used to measure the temperature of liquids, vapors, and other gas media or solid surfaces in the range of -200 ° C to 500 ° C during various manufacturing processes. It can be used in conjunction with display instruments, recording instruments, and various control systems. It has the characteristics of simple structure, easy installation and use. Currently widely used in industries such as petrochemicals, metallurgy, and power plants.

6302A aviation plug-in PT100
The resistance of PT100 platinum thermistor will change with temperature, and the temperature value can be reflected by the resistance.
Strong anti-interference ability and good long-distance transmission performance in output
The signal converter is encapsulated in epoxy resin as a module, which has the advantages of vibration resistance, corrosion resistance, moisture resistance, etc. Can be used for working conditions with poor environmental conditions
High precision, anti-interference, good long-term temperature performance, maintenance free, and capable of remote transmission (up to 1000 meters)
| Technical Specifications | |
|---|---|
| measurement range | '-200~500℃ |
| Accuracy | ± 0.25% (typical) ± 0.5% (typical) |
| long-term stability | ± 1.0% F.S/year |
| Operating Temperature | '-40℃~85℃ |
| ambient humidity | 0-95%RH, non-condensing |
| power supply | 15~30VDC |
| output signal | PT100 |
| threaded interface | G1/4&G1/2&G3/8 or user selectable |
| Shell material | Stainless Steel 304 |
| Shell protection level | IP65 |
| measurement range | PT100 thermistor: -200~0 ° C, -100~0 ° C, -50~0 ° C, -50~50 ° C; 0~50 ° C; 0~80 ° C; 0~100 ° C; 0~120 ° C, 0~150 ° C, 0~200 ° C, 0~450 ° C, 0~500 ° C or any range specified by the customer |
| load resistor | 0~850Ω |
| Cold end compensation error | 0.5%/10°C |
| Protective tube material | 304 stainless steel |
| Environmental temperature impact | 0.25%/10°C |

| Selection code | ||||||
|---|---|---|---|---|---|---|
| 6302A- | 1 | 2 | 3 | 4 | 5 | 6 |
| output signal | ||||||
| 2 | P1 | PT100 resistance signal | ||||
| 3 | Interface thread (process connection metric thread) | |||||
| J1 | G1/2 external thread | |||||
| J2 | G1/4 external thread | |||||
| J3 | G3/8 external thread | |||||
| J4 | M27 × 2 external thread | |||||
| J5 | M16 × 2 external thread | |||||
| J6 | Active flange type/fixed flange type | |||||
| 4 | Insertion depth (including threads or flanges) | |||||
| L | Indicate the actual insertion depth | |||||
| 5 | Probe diameter | |||||
| 6 | 6mm (quick delivery) | |||||
| 8 | 8mm | |||||
| X | Customer specified, technical negotiation before placing an order | |||||
| 6 | Measurement range (see the commonly used range code table below for details) | |||||
| /6 | Example: Measurement range: 0~100 ℃ | |||||
| 7 | OEM customization | |||||
| / | Regular shipment without customized content | |||||
| A | Nameless light plate customized product | |||||
| B | Designated nameplate, designated trademark product | |||||