| GPR-1100 | GPR-1000 | GPR-2000 | GPR-3500 MO | |
| measurement range | 0-10、0-100、0-1000 ppm、0-1%、0-25% | 0-1000 ppm、0-1%、0-25% | 0-1%, 0-5%, 0-10%, 0-25% | 0-100% |
| accuracy | Under constant conditions, less than 2% of the selected range | Under constant conditions, less than 2% of the selected range | Under constant conditions, less than 2% of the selected range | Under constant conditions, less than 2% of the selected range |
| response time | T90<10 seconds | T90<10 seconds | T90<10 seconds | T90<13 seconds |
| recovery time | Under nitrogen purging, reach<10 ppm within 60 seconds and<1 hour in air | Under nitrogen purging, the air reaches<100 ppm within 60 seconds and<15 minutes | Not applicable | Not applicable |
| Sensitivity (low-density lipoprotein) | 0.05ppm | 5 ppm | 0.01% | 0.10% |
| linearity | <1% of the proportion | <1% of the proportion | <1% of the proportion | <1% of the proportion |
| Sensor Model |
GPR-12-333 XLT-12-333 is suitable for gas mixtures with CO2>0.5% |
GPR-12-100-M XLT-12-100-M is suitable for gas mixtures with CO2>0.5% |
GPR-11-32-RTS XLT-11-24-RTS is suitable for gas mixtures with CO2>0.5% |
GPR-11-120-RTS |
| Sensor lifespan at 25 ° C (77 ° F) and 1 atm | Can be stored for 24 months in<1000 ppm O2 |
GPR-11-32-RTS in the air for 32 months XLT-11-24-RTS 24 months in the air |
GPR-11-32-RTS in the air for 32 months XLT-11-24-RTS 24 months in the air |
24 months in 100% O2 |
| Calibration interval | 30 days | 30 days | 30 days | 30 days |
| inlet pressure | 0.34 - 2 barg (5-30 psig), Equipped with atmospheric emission outlet | 0.34 - 2 barg (5-30 psig), Equipped with atmospheric emission outlet | 0.34 - 2 barg (5-30 psig), Equipped with atmospheric emission outlet | 0.34 - 2 barg (5-30 psig), Equipped with atmospheric emission outlet |
| traffic | 0.5-1.0 liters per meter (1-2 SCFH) | 0.5-1.0 liters per meter (1-2 SCFH) | 0.5-1.0 liters per meter (1-2 SCFH) | 0.5-1.0 liters per meter (1-2 SCFH) |
| Gas connection | 1/8 'NPT external thread quick connector | 1/8 'NPT quick connector | 1/8 'compression pipe joint | 1/4 'compression pipe joint |
| Liquid receiving components | stainless steel | Brass connection; Stainless steel optional |
GP version: brass, plastic ATEX version: Stainless steel flow casing, piping, and fittings |
stainless steel |
| display | Graphic LCD 7 x 3.5 cm (2.75 x 1.375 inches) | Graphic LCD 7 x 3.5 cm (2.75 x 1.375 inches) | Graphic LCD 7 x 3.5 cm (2.75 x 1.375 inches) | Graphic LCD 7 x 3.5 cm (2.75 x 1.375 inches) |
| solve | 0.01ppm | 1ppm | 0% | 0.10% |
| shell |
Painted aluminum NEMA 4X, 10.1 x 22.9 x 7.6 centimeters (4 x 9 x 3 inches) 3.64 kilograms (8 pounds) |
Painted aluminum NEMA 4X, 10.1 x 22.9 x 7.6 centimeters (4 x 9 x 3 inches) 3.64 kilograms (8 pounds) |
Painted aluminum NEMA 4X, 10.1 x 22.9 x 7.6 centimeters (4 x 9 x 3 inches) 3.64 kilograms (8 pounds) |
Painted aluminum NEMA 4X, 10.1 x 22.9 x 7.6 centimeters (4 x 9 x 3 inches) 3.64 kilograms (8 pounds) |
| compensation | Air pressure and temperature | Air pressure and temperature | Air pressure and temperature | Air pressure and temperature |
| signal output | 0-1V | 0-1V | 0-1V | 0-1V |
| LED indicator light | Low battery (72 hour warning); Charging mode | Low battery (72 hour warning); Charging mode | Low battery (72 hour warning); Charging mode | Low battery (72 hour warning); Charging mode |
| Operating Temperature |
5 ° C to 45 ° C (41 ° F to 113 ° F) (GPR sensor), -10 ° to 45 ° C (14 ° F to 113 ° F) (XLT sensor) |
5 ° C to 45 ° C (41 ° F to 113 ° F) (GPR sensor), -10 ° to 45 ° C (14 ° F to 113 ° F) (XLT sensor) |
5 ° C to 45 ° C (41 ° F to 113 ° F) (GPR sensor), -10 ° to 45 ° C (14 ° F to 113 ° F) (XLT sensor) |
5 ° C to 45 ° C (41 ° F to 113 ° F) |
| power | Rechargeable battery (lead-acid) | Rechargeable battery (lead-acid) | Rechargeable battery (lead-acid) | Rechargeable battery (lead-acid) |
| battery life | Up to 30 days of use on a single charge (40 hours of use when the pump is running) | Up to 30 days of use on a single charge (40 hours of use when the pump is running) | Up to 30 days of use on a single charge (40 hours of use when the pump is running) | Up to 30 days of use on a single charge |
| Regional classification |
ATEX:Ex II 2 G Ex ib IIC T4 Gb Tamb : 5 ° C to+45 ° C |
ATEX:Ex II 2 G Ex ib IIC T4 Gb Tamb : 5 ° C to+45 ° C |
ATEX:Ex II 2 G Ex ib IIC T4 Gb Tamb : 5 ° C to+45 ° C |
For general use only |