Welcome Customer !

Membership

instrumentb2bThe functions and characteristics of KDW330 handheld thermometer
1. Purpose
The KDW330 portable melting thermometer is a high-precision instrument developed specifically for the rapid measurement of molten metal temperature during the melting process in industries such as smelting and casting.
This instrument is based on a high-performance single-chip microcomputer, uses high-precision measurement sensor devices, and adopts various intelligent processing technologies to ensure long-term stable operation of the instrument. Improved the resistance of the instrument.
This instrument, when paired with suitable thermocouples, can perform precise and rapid temperature measurement in various melting scenarios.
thermocouple
model
Measurement range (℃)
Suitable occasions
Single platinum rhodium
KS-602
0~1750
Liquid steel, iron, and copper
Single platinum rhodium
KR-602
0~1750
Liquid steel, iron, and copper
Double platinum rhodium
KB-602
500~1800
High temperature molten steel
Tungsten rhenium
KW-602
0~2000
Steel and molten iron
Nickel chromium nickel silicon
K
0~1000
Aluminum and zinc liquid
2、 Characteristics
The KDW330 portable melting thermometer adopts an integrated design of thermometer and temperature gun. Based on the harsh working environment and high temperature of the melting temperature measurement site, it adopts a shock resistant and high brightness LED digital display and a fully metal shell design, making it more durable.
This instrument has:
1. The temperature value is automatically maintained during temperature measurement.
2. The function of synchronously indicating the end of temperature measurement (lifting the test gun) through lighting and buzzing sounds.
3. Alarm functions such as disconnection, over range, and power undervoltage.
4. The power supply adopts nickel hydrogen rechargeable batteries (which can generally measure temperature thousands of times after one charge).
5. Turn on the power and automatically disconnect the power function after 2 minutes of no signal input.
This instrument can be equipped with 5 types of thermocouples on the same machine.
Due to the adoption of anti-interference measures, this instrument has * * resistance. When the intermediate frequency furnace is powered on for melting, temperature measurement can be carried out inside the furnace without cutting off the power.
3、 Technical performance indicators:
1. Measurement range and measurement accuracy:
Thermocouple division number
Measurement range (℃)
accuracy class
S
0~1750
≤±0.2%±1℃
R
0~1750
WRe3/25
0~2000
B
500~1800
K
0~1000


2. Resolution (℃): 1
3. Response time (S): 1
4. Power supply: rechargeable battery DC6V ± 10%

4、 Installation and operation

  1. Installation of test gun:
    Open the packaging box, take out three sections of the test gun, hold the fixed copper head section with your hand, and rotate them one by one to avoid twisting the compensating wire and breaking the welding. After tightening, there should be no shaking sensation.

Carefully connect the two core plug inside the gun barrel to the same type of two core plug inside the connection sleeve on the portable melting thermometer (make sure to plug the signal plug into the socket completely), then insert the test gun into the connection sleeve and tighten it with a stop screw.
2. Operation of portable melting thermometer
Press the power switch (ON/OFF) on the instrument panel, and the display immediately shows "", followed by the display of the required thermocouple model (/S type,/R type,/B type,/WRe3/25,/K type)
When the thermocouple is not plugged in, the instrument displays "". When the thermocouple is plugged in and in good contact, the instrument displays the ambient temperature (the instrument with the B division mark displays ""), and temperature measurement can be performed at this time. Insert the thermocouple into the molten metal at a depth of ≥ 20mm during temperature measurement, ensuring fast, accurate, and stable operation. When the indicator light (H) is on (with sound and light prompts), the thermometer should be immediately lifted (usually for 3-5 seconds). The instrument automatically displays and maintains the measured temperature value.
Replace the used thermocouple and insert a new one (even if it is not damaged, it should be removed, paused, and reinserted into the test gun) to perform temperature measurement again.
The instrument displays "-1", indicating that the thermocouple is not plugged in or not properly inserted. Gently push or rotate the paper tube to ensure good contact.

The instrument displays "-000" and the input signal is reversed.

The instrument display shows that the input signal of "1" exceeds the range.

The instrument display "HELP" indicates that the battery is under voltage.

After the instrument measurement is completed, it should be turned off in a timely manner to avoid battery waste.

5、 Accessory kit
1. Test gun: The test gun model must match the instrument model, with specifications ranging from 1.5 to 2.5 meters.
2. Charger
3. Quick puppet head

Latest News