What are the common faults and troubleshooting methods of mortar shrinkage curing box?
Date: 2025-12-18Read: 3

1、 What are the common faults and troubleshooting methods of mortar shrinkage curing box? Overview:

The mortar shrinkage curing box is developed, designed, and manufactured according to the technical materials provided by the national building materials research and relevant departments. It is suitable for large, medium, and small cement plants, construction, water conservancy, highway and bridge engineering, construction departments, scientific research departments, and colleges and universities to perform standard curing on the strength and setting time of concrete, cement products, insulation mortar insulation materials, and building material samples.

This device has undergone a series of improvements in terms of overall design and internal structure based on the original design. The box is sealed with magnetic adhesive strips, and the outer shell and stainless steel inner liner are electrostatically sprayed. There is heating and cooling inside, which has a series of advantages such as beautiful appearance, high precision in temperature and humidity control, and durability.

2、 Technical indicators:

Instrument accuracy: ± 0.1 ℃

Temperature difference inside the box: 20 ℃ -40 ℃ ± 2 ℃ (adjustable)

Power supply voltage:~220V

Heating power: 200W

Cooling power: 150W

Humidification power: 55W

Dehumidification power: 150W


IIIWhat are the common faults and troubleshooting methods of mortar shrinkage curing box? Structure and working principle:


1. Structure:

The box is composed of three layers: inner, middle, and outer. The outer shell is treated with iron plate spraying, the inner liner is welded with mirror stainless steel, the middle layer is filled with high-quality insulation material, and the shelf is made of high-quality stainless steel pipes. The structure is sturdy, the shape is beautiful, and the corrosion resistance is good. The shelves and wall panels are interchangeable, and the automatic control system uses high-precision control instruments to cool and heat according to the given temperature and humidity. The temperature inside the box is displayed digitally by the instrument, with intuitive readings and convenient temperature adjustment.

2. Working principle:

a、 Temperature control: When the temperature inside the box is lower than the lower limit set value of the temperature controller, the temperature controller outputs a heating signal to command the electric heating element to start working. Conversely, when the temperature is higher than the upper limit set value, the temperature controller commands the refrigeration system to work. When the temperature inside the box reaches the required value, it automatically stops heating or cooling. (Note: To avoid frequent operation of the thermostat relay, the hysteresis value can be adjusted to vary the relay's operating range within 0-1 ℃.)

4、 Usage 1: After the box is in place and stationary for 24 hours, connect the ground wire and adjust the level. There is no need to add water to the water tank, and the heating tube is a new dry burning type.

3. Inspection of temperature and humidity sensors: Open the box door, hang the sensor in the middle position inside the box, and pay attention to keeping it clean and avoiding oil stains during operation.

4. Connect the power supply, check the temperature and humidity settings of the temperature and humidity controller, and start the instrument to work. When the temperature and humidity inside the box reach the set value, put the specimen in. Heating and cooling are strictly prohibited from being carried out simultaneously.

5、 Attention:

Please read the instruction manual carefully before use

1. The box should be used in a ventilated, dry, flat and non corrosive environment.

2. After the box is in place, it needs to be stationary for 24 hours before it can be powered on. When the power supply voltage is unstable, a voltage regulator must be used. After the compressor stops working, it must be restarted at least 5 minutes to ensure the normal service life of the compressor.

3. The equipment must be firmly grounded and undergo insulation testing before operation to avoid the risk of electrical leakage.

4. The tilt angle during transportation should not exceed 45 degrees, and inversion is not allowed to avoid damaging the compressor refrigeration system.

5. This device is used indoors with a working environment temperature of 0-30 ℃. We are not responsible for any violations!


7What are the common faults and troubleshooting methods of mortar shrinkage curing box? Common faults and troubleshooting methods:


1. No display on the instrument panel:

(1) Check if the power switch is damaged, and replace it with a new one if it is damaged.

(2) Check if the instrument power supply incoming line has come loose and tighten the screws.

2. The whole machine is not cooled:

(1) Is the set value of the temperature controller correct? The temperature controller is not outputting a signal.

(2) The temperature controller outputs a signal but the contactor does not operate. Replace the contactor.

(3) Check for refrigerant leakage. If refrigerant leakage occurs, please supplement F12/F406.

(4) The compressor is not working. Connect the compressor directly to the power supply. If it still does not work, it means that the thermal relay or compressor is burnt out and must be replaced.

3. The whole machine is not heated

(1) The temperature controller and contactor are burnt out and need to be replaced.

(2) The heating tube is burnt out, replace it with an electric heating element of the same power.

5. The magnetic adhesive strip is not tight enough

Use heating method to widen the magnetic tape (hair dryer can be used).

8、 Random attachments and technical documents:

One set of shelves, totaling 4 pieces

1 copy of user manual

1 copy of product qualification certificate