The main method for measuring the moisture content of water-based rock debris is through drying and weight reduction. The specific operation can be divided into two categories: specialized instrument method and traditional laboratory method. It is necessary to ensure that the measurement accuracy meets the requirements of industrial standards (usually controlled within12%Within).
1、 Special moisture analyzer method
Instrument modelAs follows:BOS-1205MAHalogen moisture analyzer, with core components made in GermanyHBMWeighing sensor, weighing resolution0.001gTemperature regulation range45℃~200℃.1
operating steps:
Calibration and zeroing: Install sample disk (diameter)90mm)And the bracket, calibrate the instrument according to the instructions and reset it to zero.
Sampling and weighing:取3gSpread representative water-based rock debris samples flat on the sample tray and record the initial weight.
Heating and drying:
o Select automatic/Manual program, set temperature (recommended)105℃~110℃)And heating mode (standard or fast).
o After startup, the annular halogen lamp heats the sample to a constant weight (the instrument automatically determines the endpoint).
Result reading:
o The instrument directly displays data such as moisture content, solid content, and moisture regain, and automatically calculates the average value.
advantage:
Automated measurement to avoid human error, with a single test approximately10~30minute;
Support data storage andRS232Interface export.
2、 Traditional drying method (laboratory standard)
Applicable scenariosWhen no dedicated instruments are available, an oven, dryer, and analytical balance are required.
step:
sampling:取50gThe above natural water bearing rock fragments can be quickly weighed(m₀), accuracy0.01g.
dry:
o Sample placement105℃~110℃Oven drying24Hour;
o Transfer to a dryer and cool to room temperature.
Redundancy and Calculation:
o Weigh the quality after drying(mₛ);
o Moisture content formula:
w=m0−msms×100%w=msm0−ms×100%
Keep the result to two decimal places.
critical control point:
The drying time needs to ensure constant weight (quality change)≤0.01g);
The cooling process must strictly prevent moisture absorption.
3、 Precautions
Sample representativenessThe rock fragments need to be mixed evenly to avoid clumping and affecting measurement accuracy;
Temperature setting: Exceeding110℃May lead to organic matter decomposition, below105℃Is it dry;
industrial standardThe moisture content of processed water-based rock debris needs to be≤15%Only then can transportation or reuse requirements be met.
Appendix: Definition of Moisture Content
Wet basis moisture content:M=100×(wet weight−dry weight)/wet weightM=100×(wet weight−dry weight)/Wet weight(≤100%);
Dry basis moisture content:M=100×(wet weight−dry weight)/dry weightM=100×(wet weight−dry weight)/Dry weight (can be greater than)100%).