Welcome Customer !

Membership

Help

Veolia Sievers analyzer
Custom manufacturer

Main Products:

instrumentb2b>Article

Veolia Sievers analyzer

  • E-mail

    sievers.china@veolia.com

  • Phone

  • Address

    Building 5-6, Chuangqi Tiandi, 1761 Zhangdong Road, Pudong

Contact Now
Successful correlation between BOD and TOC in ethylene wastewater process control
Date: 2025-09-22Read: 0

challenge

An ethylene production plant is seeking to improve the performance and monitoring of wastewater treatment processes. After the wastewater from the production equipment converges at the lifting station, it flows into the homogenization tank (EQ tank). Before the wastewater enters the Dissolved Air Flotation (DAF) system, operators add treatment chemicals to the water and adjust its pH value. The treated water is sent to the biological treatment system for further processing, and then clarified and discharged.


The factory needs to take samples at the discharge outlet every day and analyze them using Biochemical Oxygen Demand (BOD). BOD and other measurement data of water are used for compliance testingTry to calculate the total amount of organic matter emitted by the factory. However, the factory is unable to use the low BOD value reported as' Non Detect '. Another challenge is that BOD analysis requires a reporting time of 5 days, which makes it impractical to apply BOD analysis to monitoring and optimizing treatment processes.

Solution


The factory adoptsAnalyze monitoring planTo optimize water treatment processes to reduce organic emissions. Although the time requirement of BOD analysis makes this analysis method lose practical application value, the relationship between BOD and TOC can be used to establish the correlation between the two at each sampling point. Using these correlation coefficients for Total Organic Carbon (TOC) analysis, report almost real-time monitoring data, and predict the relevant biochemical oxygen demand (BOD) within minutesC,Biochemical Oxygen Demand-Correlated)” 数据。


When establishing correlation, a sampling plan is needed to define sample collection and data analysis. At the operation site, the factory usesSievers®InnovOx Laboratory TOC AnalyzerReport the initial data of relevance. After successfully establishing the correlation, the factory can switch the analysis mode to online analysis at any time.

乙烯废水工艺控制中成功关联BOD和TOC

From left to right:Sievers InnovOx

Online and laboratory TOC analyzer


The factory selects three sampling points to determine process operations and compares TOC and BOD data (see Table 1).

Sampling point location

Process operation

Upgrade Station

Flow towards the homogenization tank and adjust the pH value

Final clarification pool

Reduce total suspended solids (TSS)

Deep processing pool

none

Table 1: Wastewater Sampling Points


Due to the nonlinearity of BOD analysis data, it is required to separately derive the correlation coefficients of BOD and TOC samples at each sampling point.


Sampling multiple times a day can improve the accuracy of correlation. In this study, the factory monitored three testing points and extracted a total of seven samples within two weeks.


The first sampling point is located after the lifting station and before the homogenization tank, and the sample taken comes from a stable inflow. The measurement data is listed in Table 2.

Stable inflow


_

TOC

BOD

BODC

May 30th

44.7

90.5

109.07

May 31st

50.7

124

118.54

June 1

45.3

104.4

110.02

June 6

40.7

84.8

102.76

June 7

25

101

77.99

June 8

24.9

82.4

77.83

June 13

93.4

195

185.91

Table 2: Inlet data of homogenization tank


Any obvious outliers are replaced by the average value of BOD before and after, thereby increasing the correlation coefficient from 0.675 to 0.923. For wastewater, correlation coefficients higher than 0.5 can be used. The BODC values in Table 2 are calculated based on the relationship between measured BOD and TOC values.


BOD/TOC correlation

乙烯废水工艺控制中成功关联BOD和TOC

Figure 1: Correlation between BOD and TOC in the influent of the homogenization tank


The correlation between BOD and TOC in the inflow is very reliable, so TOC can be used as a substitute for BOD (see Figure 1).


The measurement results of the second sampling point at the outlet of the final clarification tank show that if the concentration is too low, the BOD value cannot be determined (see Table 3).

Water outlet


_

TOC

BOD

May 30th

5.97

3.79

May 31st

6.85

3.39

June 1

8.19

3.95

June 6

1.08

Not detected

June 7

7.33

3.83

June 8

7.7

Not detected

June 12

6.76

4.23

Table 3: Final Clarifying Pool Data


Although the carbon content measured by TOC analysis has changed by 8 times, the sensitivity of BOD still does not meet the requirements for quantitative data. The BOD data in Table 3 shows that out of 7 samples, 2 samples could not be quantified and were reported as' undetected '. The data deviation between the other 5 BOD samples is within+/-4%, making it difficult to distinguish statistically. The BOD of the effluent can only be used for qualified/unqualified testing.


The BOD data of the deep processing pool (see Table 4) were reported as' undetected ', therefore it is not possible to establish a correlation with TOC. Although BOD is reported as' undetected ', TOC data is still accurate, precise, and linear.

Deep processing pool


_

TOC

BOD

May 30th

6.94

Not detected

May 31st

7.57

Not detected

June 1

8.45

Not detected

June 6

7.85

Not detected

June 7

6.72

Not detected

June 8

6.11

Not detected

June 12

6.79

Not detected

Table 4: Deep Processing Pool Data

Conclusion

This ethylene production plant has successfully utilizedTOC analysis methodTo monitor the wastewater treatment process. The inflow correlation coefficient they obtained is very reliable, so the almost real-time TOC analysis method can be used instead of the commonly used 5-day BOD testing method.


The organic carbon measurement results are the most reliable wastewater discharge data. The TOC analysis method can directly measure low ppm organic carbon in water, making it a more reliable monitoring and optimization tool. Operators can quickly respond to potential issues based on real-time data and take corrective measures in a timely manner.