What are the operating procedures for an infrared oil analyzer?
Date: 2017-11-10Read: 0

Infrared oil analyzerHow to use it normally and what are the operating procedures for the infrared oil analyzer? The article provides a detailed explanation for you.

What are the operating procedures for an infrared oil analyzer?

The detailed operating procedures for the infrared oil analyzer are as follows:

1. Operation steps of infrared oil analyzer:
1) Turn on the power (oil meter power);
2) Enter the software;
3) Measurement target (measuring in water bodies)

The detailed operating procedures for the infrared oil analyzer are as follows:

1. Operation steps of infrared oil analyzer:

1) Turn on the power (oil meter power);

2) Enter the software;

3) Measurement target (measuring oil in water);

4) Adjust the fullness (open the cassette lid and observe that the fullness is zero);

5) Place the colorimetric dish containing carbon tetrachloride into the cassette, adjust the fullness to between 80% and 85%, and confirm;

6) Click on regular measurement to establish a platform (automatically measure blank spectra);

7) Replace the extracted sample (input sample, volume of extractant), click on wide range measurement;

8) Obtain concentration results and print them;

9) Exit the software and turn off the power of the oil meter;

2Precautions for operating infrared oil analyzer

1) When the infrared oil analyzer is fully adjusted to normal (wavelength around 3.28 microns), there is a yellow green spot in the middle of the colorimetric cell that is a normal spot;

2) When the instrument wavelength suddenly becomes unstable, the host power should be cut off in a timely manner and the computer should be restarted with a cold start;

3) Pay attention to regularly maintaining the computer and backing up important monitoring data. To avoid being damaged by viruses and causing losses. Regularly clean up computer viruses and protect your software;

4) Avoid direct sunlight and ensure good indoor ventilation;

5) Note that if the purity of carbon tetrachloride is not sufficient, please refine it before use;

6) Pay attention to the cleanliness of the colorimetric dish and the direction of the arrow on the colorimetric dish;

7) When measuring water samples, pay attention to the dehydration of the water sample and do not forget the extraction ratio;

8) When storing the instrument, it should be kept in a clean room with a temperature requirement of 0-35 degrees Celsius, a humidity of less than 85%, and a non corrosive environment.

Energy Spectrum Technology is a diversified high-tech enterprise dedicated to the research, development, production, and sales of Fourier transform infrared spectrometers, infrared oil analyzers, and dust free silica analyzers; Whether for routine examinations or for cutting-edge scientific research, you will definitely find the ideal tool that suits you here.