Agilent Network Analyzer E8362C Maintenance Case Sharing
Date: 2019-06-12Read: 0

A unit in Xi'an sent an Agilent network analyzer E8362C for repair, and the feedback fault was that it could not enter the testing interface normally when turned on and remained on the Windows desktop. Let's take a look at how Mr. Li from Antai Testing and Maintenance Center repaired it together:

1、 Instrument model

Agilent Network Analyzer E8362C.

2、 Fault phenomenon:

Unable to enter the testing interface properly after booting up, stuck on the Windows desktop.

3、 Detailed explanation for disassembly and maintenance:

Testing process:

After powering on the instrument, it stayed on the Windows desktop and upon inspection, it was found that there were no network distribution programs or calibration files.

Next, first, backup the original system. Then take measures to restore the system, and the instrument prompts Phaselock loss.

Because the instrument still cannot boot up and display normally after reinstalling the system, the next step is to disassemble and inspect whether there are any burnt parts inside. After dismantling and testing, it was found that the microwave switch of the instrument was damaged, and the internal micro components were burnt with large signals, causing bubbles and resulting in low output power at port 1, while port 2 was normal. It is preliminarily judged that the switch is faulty.

Maintenance measures:

After discovering the current fault issue, replacement and repair measures were taken: replacing the damaged micro components of the microwave switch, adjusting the instrument parameters of the whole machine, and the startup test still did not pass. Due to severe internal burning damage of the instrument, replacing the micro components still cannot solve the problem. As the switch components are severely burned and cannot be repaired, the microwave switch must be replaced as a whole. After replacing the microwave switch, the entire machine was adjusted and the testing instrument was turned on and tested normally.

Repair Summary:

Based on the experience of repairing the instrument, the inability to access the testing interface can sometimes be solved not only by reinstalling the system, but also by severe internal burning of the instrument. The main reason for this phenomenon is also due to the serious burning of the instrument caused by improper use.

Precautions for using RF instruments:

1. Pay attention to electrostatic protection, especially the electrostatic protection of exposed interfaces;

2. Be careful to avoid hot plugging of interfaces: connect the interface first, and then add the signal; Disconnect the signal first, then disconnect the interface connection;

3. Regularly maintain and upkeep the instruments.

Antai test repairable faults:

Antai test repairable faults:

1. Power failure: unable to power on normally

2. Input end fault: The port is not responding properly; No output; No input; Signal anomaly

3. Test analysis failure: abnormal test results; Calibration failed; extremely poor

4. Display malfunction: flower screen; black screen

5. Button malfunction: No response from the button; The adjustment knob is unresponsive

6. Interface malfunction: Storage medium not recognized; Cannot connect with the control system

7. Other usage issues, etc.

Shenzhen Meijiate Technology Co., Ltd. provides long-term supply of:
Spectral analyzer/wavelength meter: Yokogawa AQ6370,Yokogawa AQ6370B,Yokogawa AQ6370D,Anritsu MS9740A,KEYSIGHT 86120B/C ,KEYSIGHT 81622A/B/C
Network analyzer: KEYSIGHT E5071C, KEYSIGHT E5080A, KEYSIGHT E5080B, R&S ZNB8, R&S ZND8
Spectrum analyzer: KEYSIGHT N9040B, KEYSIGHT N9030B, R&S FSW8, R&S FSW26
Optical attenuator: KEYSIGHT 81576A,KEYSIGHT 81560A,KEYSIGHT 81570A,KEYSIGHT N7764A,KEYSIGHT N7762A,KEYSIGHT N7761A
Optical oscilloscope: KEYSIGHT 86100D,KEYSIGHT N1092A,KEYSIGHT N1092D,KEYSIGHT N1092E,KEYSIGHT 86107A,KEYSIGHT 86116C 025,KEYSIGHT 86105D 281
Adjustable light sources: KEYSIGHT 81600B, KEYSIGHT 81640A, KEYSIGHT 81680A, SANTEC N7714A
Optical power meter probe/optical power host: KEYSIGHT 8164B,KEYSIGHT 8163B,KEYSIGHT 81628B,KEYSIGHT 81626B,KEYSIGHT 81624B