Common faults and solutions of PH instruments
Common faults and solutions of PH instruments
Phenomenon |
reason |
resolvent |
Controller has no display |
The power supply is not connected or the fuse is blown |
Check the power supply circuit and fuse |
Display numbers but jump up and down |
The electrode wire is damp or short circuited
Electrode ceramics contaminated
|
Check the electrode wires
Cleaning ceramic holes (0.1N HCL)
|
The standard solution pH 4.01 and pH 6.86 cannot be adjusted |
External contamination of electrodes
The electrode wire is damp
Electrode damage or cracks
Electrode ceramic blockage
|
Cleaning electrode (0.1N HCL)
Check the electrode wires
Replace the electrode
Cleaning ceramic holes (0.1N HCL)
|
Slow response to numbers |
Electrode film or ceramic hole contaminated |
Clean the entire electrode |
|
When the actual pH value has changed significantly and the controller value
But the change is very small
|
Electrode ceramic hole blockage
Electrode aging, slow response
|
Cleaning electrode (0.1N HCL)
Use regeneration solution or replace the electrode
|
Control the display number to remain unchanged |
Electrode rupture |
Replace the electrode |
The relay in the controller does not operate or cannot add medication |
Relay damaged
Controller setting error
Insufficient medicinal solution
|
Supplier replaces relay
change setting
Supplement medication solution
|