E-mail
654388291@qq.com
-
Phone
18610205680
-
Address
Room 257, Q Zone, 2nd Floor, Building 1, No. 9565 Huqingping Road, Qingpu District, Shanghai
Shanghai Yingle Instrument Co., Ltd
654388291@qq.com
18610205680
Room 257, Q Zone, 2nd Floor, Building 1, No. 9565 Huqingping Road, Qingpu District, Shanghai
04.25,20250
The medical suture diameter tester, also known as the suture diameter measuring instrument or suture diameter detector in the industry, is developed and produced by Shanghai Yingle manufacturer. It is designed according to YY1116-2020 and YY0167-2020 standards and is specifically used for measuring suture diameter. The instrument is equipped with a Linux intelligent controller, force measuring mechanism, distance measuring mechanism, micrometer, printer, and other components. Chinese and English settings, touch screen control, no training required, one click intelligent testing. Nominal specification: Suitable for all medical sutures.
Technical Indicators:
Operating interface: Simplified Chinese/English
Test standard: YY1116-2020 "Absorbable surgical sutures"
YY0167-2020 Non absorbable surgical sutures
Display: Adopting a 7-inch K600+core/dot matrix 65 color 800 * 480 resolution touch screen display, the PLC control system dynamically displays the loading force value; Optional operating interface: Chinese/English/switching between Chinese and English; Digital closed-loop control, AD acquisition speed of 500KHz, internal code of 500000; High precision clock chip, power-off storage; Overload alarm shutdown protection; Touch operation: nominal specifications, test load, printing, testing, left row, right row, time, calibration. Weight loading tension, automatic recording of wire diameter data, automatic output of data report, and printing of test results by onboard printer.
Test data: The onboard system automatically collects measurement data at 3 points 0 degrees and 90 degrees (line 1/4, 1/2, 3/4 positions) and calculates the average value. It can calculate the average value, total average value, and deviation value, and automatically output a data report.
Programmable controller: PLC+ARM,
Sensor: High precision force sensor with software and hardware protection function
Nominal specifications: 12-0... 10 for all types of medical sutures.
Measuring range: 0-5mm, error not greater than ± 1 ‰, resolution: 0.001mm
Pressure foot diameter: 12.7 ± 0.02 mm; reference surface diameter: 50 mm.
Flatness between the presser foot and the reference surface: not greater than ± 0.005mm.
Parallelism between the presser foot and the reference plane: not greater than 1 ‰
Loading weight: 0-2Kg, accuracy: ± 0.1g
Foot load: ≤ 60g, 210g ± 3g.
Printing method: Airborne micro needle printer
Power supply: AC90V-240V/50Hz
Net weight: 38kg Dimensions: 320 × 1200 × 280 (mm)
Last Video: Disposable sterile injection needle toughness tester