WS4320 Industrial Online Primary Battery Dissolved Oxygen Electrode
WS4320 Industrial Online Primary Battery Dissolved Oxygen Electrode |
1、 WS4320 Industrial Online Primary Battery Dissolved Oxygen Electrode |
|
2、 WS4320 Industrial Online Primary Battery Dissolved Oxygen Electrode Technical Parameters |
measurement range |
0.00-40.00 or 0.0-400.0% (gel electrolyte) |
accuracy |
± 2% of full range (25 ℃) |
repetitiveness |
± 2% of full range (25 ℃) |
Zero point output |
Less than 0.2mV (25 ℃) |
Saturated output |
25 ± 5mV (in air or saturated water, 25 ℃) |
response speed |
30S |
Main material |
POM or Sai Steel Pipe |
temperature compensation |
Thermistor compensation 0-50 ℃ |
Electrode size |
12*130mm |
Cable length |
6 meters or 10 meters |
electrode terminal |
Needle shaped 4-core connector |
|
3、 Application |
|
Scope of application: wastewater, aquaculture, chemical industry, seawater and other fields.
Precautions for use: 1. Remove the protective cap or sheath at the bottom of the electrode and rinse the electrode with pure water. Then dry off the moisture and connect the electrode to the instrument Adjust the instrument in percentage display mode, check electrode stability, and ensure repeatability is within ± 2%. 2. The electrode needs to be soaked in water for 20 minutes before its performance can be restored. 3. Electrode cleaning: Generally, soak in 3% dilute hydrochloric acid for 10 minutes. 4. The electrode film is made of plastic film. Please make sure not to use sharp objects to scratch or damage the sensor.
|