DSVDilute solution viscosity meter/Y501 relative viscosity meterIt is the earliest to provide comprehensive services to customersViscosity testing system. Whether you need a manual detection system for limited sample size testing or a fully automated injection detection system for high-throughput, safe, and highly reproducible testing, DSV can meet your needs.


Unlike the glass capillary viscometer, which detects the time it takes for a solution to pass through the capillary,the United StatesHMJcompanyDSVThe dilute solution viscometer uses pressure changes to detect relative viscosity. Two capillaries are connected in series, and the injection valve is located on both capillaries 1 and 2 middle The differential pressure sensor detects the pressure drop at both ends of each capillary tube. When the sample is injected into the second capillary, the differential pressure sensor detects the capillary2Changes in pressure on top (Figure)3). Relative viscosity is equal to the ratio of two capillary pressure drops multiplied by the instrument constantK.



Due to the high sensitivity of the relative viscometer, it is only necessary to detect a sufficiently diluted solution to directly obtain the desired viscosity without extrapolating from the United StatesHMJcompanyDSVThe dilute solution viscosity meter provides an accurate and high-precision method for fully automatic detection of solution viscosity.





Application advantages


❖ Save more than traditional methods50%The above solvents


❖ The detection speed is much faster than traditional glass viscometers


❖ Higher accuracy and precision


❖ *Not in contact with solvents, providing a truly safe operating environment for operators


❖ Flexible configuration according to customer needs



DSVThe relative viscometer can output the following results:




Application field


➠Polyethylene terephthalate(PET)


➠polyvinyl chloride(PVC)


➠polyolefin(Polyethlene)


➠polybutadiene(PBD)


➠Polycarbonate(PC)


➠Various types of polysaccharides


➠ synthetic rubber(EPDM)


➠ polylactic acid (PLA)


➠ hyaluronic acid(HA)


➠ cellulose


➠ nylon



Technical Specifications



Y-501host


Measuring principle

Dual capillary dual pressure sensor

test scope

0 – 20 cp(mPa. s)Or1 - 10Relative viscosity

sensitivity

1.0 x 10-4 Increasing specific viscosity

Accuracy of relative viscosity testing

+/- 0.1%relative standard deviation

Typical flow velocity

1 –3 mL/min.

Typical analysis time

2 –4 min.

Typical sample volume

20–25 mLRepeat the injection twice

Capillary size

0.02 in. ID x 12 in.

shear rate(typical)

1000 1/sec

temperature range

15–150°C

output signal

0 –10 V





autosampler


software control

passETATMDilute solution viscosity control software package

Number of sample bottles

30





Software system and result output


data collection

Test time:infinite


Number of data collection points:infinite

output result

Relative viscosity(Relative Viscosity)

Increasing specific viscosity(Specific Viscosity)

Logarithmic viscosity of specific concentration(Inherent Viscosity)

Specific viscosity(Reduced Viscosity)

Intrinsic viscosity(Intrinsic Viscosity:Soloman-GottesmanTheBillmeyerThe

HugginsTheKramerTheBaker-Philopoff)

Absolute viscosity(Absolute Viscosity)

Dynamic viscosity(Kinematic Viscosity)

molecular weight