Welcome Customer !

Membership

Help

Shanghai Kelute Technology Instrument Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Solution>Solution for formaldehyde emission from artificial boards
Solution

Shanghai Kelute Technology Instrument Co., Ltd

  • E-mail

    513687859@qq.com

  • Phone

    18317070697

  • Address

    No. 888 Siyi Road, Jiading District, Shanghai

Contact Now

The operation method is to place a sample with a surface area of 1 square meter in a climate chamber where the temperature, relative humidity, air flow rate, and air replacement rate are controlled at certain values. Formaldehyde is released from the sample and mixed with the air inside the box. The air inside the box is regularly extracted and passed through an absorption bottle containing distilled water. All formaldehyde in the air is dissolved in water; Measure the amount of formaldehyde in the absorption solution and the volume of extracted air, calculate the amount of formaldehyde per cubic meter of air, expressed in milligrams per cubic meter (mg/m ³). The extraction is periodic until the formaldehyde concentration in the air inside the climate chamber reaches a stable state.

The main body of the 1 cubic meter formaldehyde test chamber adopts the internationally advanced jacket structure. The sandwich structure has a test chamber for the inner container and an insulated box for the outer layer. The inner liner material is double-sided welded with mirror 304 stainless steel material, and the welding seam in the gas chamber is polished. The inner liner has good airtightness, increases cleanliness and reduces adsorption, and also ensures no cross contamination of the measured gas. The geometric shape of the climate chamber liner is designed to have a certain curvature, which is very beneficial for reducing air stagnation, thereby reducing temperature and humidity fluctuations, and improving gas uniformity; The inner chamber of the climate chamber can be effectively cleaned with solvents and high-purity distilled water, and the pipelines, joints, sealing gaskets, etc. inside the test chamber are made of non absorbent, non-volatile and highly stable materials, ensuring the rigor of VOC testing.

model

KLT-NT02

KLT-NT1

Nominal internal volume (liters)

225(1±2%)

1000(1±2%)

internal dimensions

(mm)

W

550

850

D

600

1080

H

680

1100

external dimensions

(mm)

W

1100

1500

D

796

1300

H

1520

1850

sex

can

temperature range

15~30 ℃

humidity range

35~70 % RH

temperature deviation

≤ ±0.5 ℃

Humidity deviation

≤ ± 2% RH

Temperature Uniformity

≤±0.8 ℃

Humidity uniformity

≤ ±2 % RH

temperature fluctuation

≤ ±0.5℃

Humidity fluctuation

≤ ± 2% RH

Ventilation rate

0-1.5 times/hour

sealing performance

The air leakage rate inside the cabin is less than 5%. If the gas supply rate or pressure exceeds 1kpa, the gas leakage rate is less than 0.5% of the cabin capacity per minute

Central wind speed

0.1-1.0m/s (continuously adjustable)

Relative positive pressure

10±5pa

spike recovery rate

>80%

Background value

Formaldehyde: ≤ 0.006mg/m3

Temperature control method

Indirect temperature control method for water circulation

Humidity control method

Dual directional adjustment method for dry gas and moisture ratio

material

shell

Q235A cold-rolled steel plate 1.5mm, surface electrostatic spraying, white

inner box

SUS304 mirror stainless steel 1.5mm (8k mirror)

adiabatic

High density polyurethane foam+XPS

seal

Silicone (food grade), polytetrafluoroethylene

Operation control

monitor

7-inch color touch screen, resolution 800x480, remote control (optional)

Automatically record projects

Temperature, humidity, pressure (optional), flow rate (optional), operating time, fault record

Control method

Temperature control: proportional, integral, derivative (P.I.D); System control: PLC+HMI

resolution

Temperature: 0.1 ℃; Humidity: 0.1% RH

communication interface

1 USB-A, 1 USB-B, 1 RS232, 1 RS485, 1 RJ-45 (optional)

Rough storage and recording function

Memory, USB flash drive, SD card

Supplement water quality and cycle

Distilled water; >30 days (continuous operation)

power supply

AC380 (1 ± 10%) V (50 ± 0.5) Hz three-phase four wire+protective grounding

Installed capacity (Kw)

2.5

3.0

Weight (Kg)

280

320

Noise (dB)

≤ 60dB

System protection

High and low pressure protection, overcurrent protection of refrigeration system; Overtemperature protection; Motor overcurrent protection; Water shortage protection; Power supply under phase, phase sequence protection, etc.