What are the manufacturers of a one cubic meter VOC formaldehyde testing chamber
Date: 2018-05-02Read: 1

Nowadays, people are paying more and more attention to health. Previously, they only believed that food could affect our health, but now we know that air can also affect our health, such as formaldehyde, TVOC, and so on. At this time, many manufacturers are also seizing the current needs of people and promoting their productsOne cubic meter VOC formaldehyde testing chamberSo which manufacturer is good.

At present, there are not many manufacturers that can independently produce and sell one cubic meter VOC formaldehyde testing chambers, but our company has been researching this product for several years. However, the previous industry did not pay attention to these, so most of our customers only know that we produce constant temperature and humidity chambers, aging rooms, etc. However, now we are starting to pay attention to indoor environments. After all, the harm of formaldehyde is so great that we are also one of the first manufacturers to study the standard GB18580-2017.

One cubic meter VOC formaldehyde testing chamber is used to detect the variation pattern of VOC release in products and classify the products into grades. Commonly used for release detection of products in industries such as artificial boards, furniture, sofas, toys, automotive interiors, and electronic products.

model

SY21-N1

SY21-H1

Nominal internal volume (liters)

1000(1±2%)

internal dimensions

(mm)

W

850

D

1080

H

1100

external dimensions

(mm)

W

1530

D

1285

H

1572

sex

can

temperature range

15~40 ℃

15~250℃

humidity range

30~80 % RH

5~80 % RH

temperature deviation

≤ ±0.5℃

≤± 0.5 ℃ (at operating temperature)

Humidity deviation

≤ ± 2% RH

≤± 2% RH (at operating temperature)

Temperature Uniformity

≤±0.8 ℃

≤±0.8 ℃(At working temperature)

Humidity uniformity

≤ ±2 % RH

≤± 2% RH (at operating temperature)

temperature fluctuation

≤ ±0.5℃

≤± 0.5 ℃ (at operating temperature)

Humidity fluctuation

≤ ± 2% RH

≤± 2% RH (at operating temperature)

Ventilation rate

0-2.5Times/hour

sealing performance

The air leakage rate inside the cabin is less than 5%. If the gas supply rate or pressure exceeds 1kpa, the gas leakage rate is less than 0.5% of the cabin capacity

Central wind speed

0.1-1.0m/s(Continuously adjustable)

Relative positive pressure

10±5pa

spike recovery rate

>80%

Background value mg/m3

Formaldehyde: ≤ 0.006; Single VOC: ≤ 0.002; TVOC:≤0.02

Temperature control method

Air jacket method (cold heat balance method)

Humidity control method

Dual directional adjustment method for dry gas and moisture ratio

material

shell

Q23Cold rolled steel plate 1.5mm, surface electrostatic spraying, white

inner box

SUS304Mirror stainless steel 1.5mm (8k mirror)

adiabatic

High density polyurethane foam+XPS

Mineral fiber

seal

Silicone (food grade), polytetrafluoroethylene

Operation control

monitor

7Inch color touch screen, resolution 800x480, remote control (optional)

Automatically record projects

Temperature, humidity, pressure (optional), flow rate (optional), operating time, fault record

Control method

Temperature control: proportional, integral, derivative (P.I.D); System control: PLC+HMI

resolution

Temperature: 0.1 ℃; Humidity: 0.1% RH

communication interface

1One USB-A, one USB-B, one RS232, one RS485, and one RJ-45 (optional)

Rough storage and recording function

Memory, USB flash drive, SD card

Supplement water quality and cycle

Distilled water; >30 days (continuous operation)

power supply

AC380(1 ± 10%) V (50 ± 0.5) Hz three-phase four wire+protective grounding

Installed capacity (Kw)

3.0

7.0

Weight (Kg)

350

380

Noise (dB)

≤ 65dB

System protection

High and low pressure protection, overcurrent protection of refrigeration system; Overtemperature protection; Motor overcurrent protection; Water shortage protection; Power supply under phase, phase sequence protection, etc.