The constant temperature and humidity test chamber is a temperature and humidity experimental instrument and equipment. In the process of operating the constant temperature and humidity test chamber, we should fully recognize the hazards caused by improper operation.
1. To ensure the safety of equipment and testing, please install external protective grounding and supply power according to technical specifications;
2. The equipment is strictly prohibited from being used for testing flammable, explosive, toxic, and highly corrosive materials;
3. Non professionals are not allowed to disassemble or repair;
4. The equipment should have reliable grounding;
5. Do not open the box door or enter the box unless necessary during the experiment, as it may cause personal injury and equipment malfunction;
6. The door lock of the equipment box can only be opened from the outside, and there must be someone to supervise entering the box;
7. If a heating sample is placed inside the box, please use an external power source for the sample and do not directly use the device's own power source;
8. The equipment is equipped with multiple protective measures, please check regularly;
9. When the temperature inside the box is ≥ 55 ℃, do not turn on the refrigeration compressor to ensure its long service life and normal operation;
10. When transporting products with refrigeration function, the inclination angle should not exceed 45 ℃. After placing them in place, they should be left to stand for 1-2 days before starting up, which is beneficial for the refrigeration system to work normally and extend its service life;
11. Do not open/close the product box door by gravity, otherwise it may cause the box door to fall off, resulting in product damage and injury accidents;
12. When the product is not used for a long time, regular moisture removal treatment should be carried out to avoid damaging the relevant components;
13. When handling products, care should be taken to avoid damaging vulnerable components such as instruments on the panel;
14. Please read the attached documents carefully before operating this device;
15. This test chamber does not have explosion-proof devices and cannot be filled with flammable and explosive materials for drying
16. During operation, please do not open the box door unless necessary, as it may cause personal injury and equipment malfunction. High and low temperature steam rushing out of the box is very dangerous, and the inside of the box door still maintains high temperature, causing burns.