Welcome Customer !

Membership

Help

Shanghai Jintong Instrument Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Article

Shanghai Jintong Instrument Co., Ltd

  • E-mail

    1415268707@qq.com

  • Phone

    13764885152

  • Address

    Room 4597, Building 6, No. 1150 Lanfeng Road, Fengxian District, Shanghai

Contact Now
From Installation to Calibration: Olympus Industrial Microscope Full Process Operation Manual
Date: 2025-09-22Read: 0
Industrial microscopes are core equipment in fields such as material analysis and electronic component testingat the beginningThe standardization of usage directly affects the accuracy of subsequent testing results.Olympus Industrial MicroscopeKnown for its high stability and precision, butat the beginningStrictly follow the entire process of "installation debugging calibration" when using. The following is a detailed step-by-step guide:
1、 Installation preparation:
Firstly, it is necessary to choose an installation environment that meets the requirements: the laboratory should avoid direct sunlight and strong vibration sources (such as air conditioning outdoor units and large instruments), the room temperature is recommended to be controlled at 20-25 ℃, and the humidity should be maintained at 40% -60% (to prevent lens mold). Place the microscope body on a stable workbench (preferably a shock resistant platform), ensuring that at least 30cm of operating space is reserved around it.
Check if the random accessories are complete, including the host, power cord, light source module (LED or halogen lamp), objective lens (usually including 5 ×, 10 ×, 20 ×, 50 ×, etc.), eyepiece (10 × or 15 ×), stage fixture, filter set, and user manual. Special note: If it is a fluorescent model, it is necessary to confirm the installation position and protective measures of the mercury lamp or LED excitation light source.
2、 Hardware assembly:
1. Connect the power supply and light source: connect the power line to the power interface on the back of the microscope (pay attention to the voltage matching, such as 220V/110V), turn on the power switch, and adjust the light source brightness knob to a lower level (to avoid sudden strong light damaging the sample or eyes). For halogen lamp models, wait for about 5 minutes to reach a stable light emission state; LED light sources can be turned on and used immediately.
2. Install the objective lens and eyepiece: Rotate the objective lens converter to align the low magnification objective lens (such as 5 × or 10 ×) with the through hole (usually marked with white alignment), avoiding excessive force that may cause the objective lens to shift; Insert the eyepiece into the eyepiece tube and adjust the eye mask position according to the observer's vision (ensuring no light leakage).
3. Fixed sample: Place the sample to be tested (such as metal slices or circuit boards) on the stage and fix it with a spring clip or vacuum suction cup (depending on the type of stage), ensuring that the sample is located directly below the through-hole and the edge does not exceed the range of the stage.
3、 Basic debugging:
After turning on the light source, first adjust the coarse focusing knob and slowly lower the stage (or raise the objective lens) until the objective lens is close to the sample but not in contact (at a distance of about 1mm); Then rotate the coarse focus knob in reverse until a blurry image is observed through the eyepiece; Use the fine focus knob to fine tune until the image is clearer. If it is in bright field mode, the height of the spotlight needs to be adjusted (usually through the lifting knob) to focus the light on the surface of the sample, and the aperture size needs to be adjusted to control the amount of light entering.
If you need to switch observation modes (such as dark field, polarization), you need to install corresponding filters or accessories, and adjust the light source angle or polarizer direction through the mode switching button. For example, in dark field mode, the annular aperture needs to be matched with the objective lens to block direct light and only use scattered light for imaging, highlighting the concave and convex details of the sample surface.
4、 Key calibration:
Industrial microscopes are commonly used for size measurement or defect analysis, so they must undergo "optical calibration" and "mechanical calibration" before use.
Optical calibration (field of view calibration): Place a standard micrometer (with known scale spacing, such as 1mm/100 grids) on the stage, first use a low-power objective lens to find a clear image, and record the length of the eyepiece field of view corresponding to N grids on the micrometer; Repeat the measurement with a higher magnification objective lens and correct the reading error of the eyepiece micrometer by calculation (for example, if one grid under a 10 x objective lens is actually 0.01mm, but the eyepiece micrometer displays 0.012mm, then multiply the correction factor in subsequent measurements).
Mechanical calibration (stage movement calibration): Some models support electric stages and require software calibration of the linear accuracy of X/Y axis movement - place the standard sample on the stage and record the displacement value displayed by the software when moving 1mm. If there is a deviation (such as displaying 1.02mm), enter correction parameters in the software to ensure the accuracy of image stitching or size measurement.
5、 Closing and Maintenance: The Key to Extending Equipment Lifespan
After the initial use is completed, the objective lens should be turned back to the low magnification position (to avoid long-term pressure on the high magnification objective lens), the light source should be turned off, and the power cord should be unplugged; Gently wipe the surface of the objective lens and eyepiece with specialized lens cleaning paper (if there are stains, clean them with a mixture of ethanol-C ₄ H ₁₀ O first); When not in use for a long time, it is recommended to cover with a dust cover and remove the fixture of the loading platform to store in a dry box.
The entire process from installation to calibration is not onlyOlympus Industrial MicroscopeThe "introductory compulsory course" is the foundation for ensuring the reliability of testing data. Only by strictly regulating every step can this precision instrument play a significant role in industrial quality inspection.