Introduction:
This method is suitable for determining the residual levels of benzene derivatives and refrigerants in the air of workplaces, production workshops, and other environments.
Method principle:
Inject the gas sample automatically into a gas chromatograph equipped with a hydrogen flame ionization detector, and measure the content of each component on dedicated columns for benzene derivatives and refrigerants through selection valves.
Reagents and materials:
Standard gases: Benzene -0.0994mg/m3, Toluene -0.0968mg/m3, p-xylene -0.2030mg/m3, o-xylene -0.1000mg/m3, dichloromethane 1.0200mg/m3, pentafluoroethane-1.0400mg/m3, 1.1.1.2-Tetrafluoroethane-1.0300mg/m3, difluorochloromethane -1.0200mg/m3.
Gas source: nitrogen (purity ≥ 99.999%), hydrogen (purity ≥ 99.999%), compressed air;
Standard gas dilution gas: N2, purity ≥ 99.999%
Instruments and equipment:
Gas chromatograph: Jing and AnalyElf PG6 online version (GC-FID)
Sample injector: Two position rotary valve (1ml sample quantification ring)
Chromatography column: benzene series/refrigerant special sleeve column
Data workstation: 3097 online version.