Detection of residual epoxyethane and 2-chloroethanol in medical devices
1. Analysis purpose
Use gas chromatography to detect residual levels of ethylene oxide (EO) and 2-chloroethanol (ECH) in medical devices.
2. Experimental instruments and standard samples
G51Pro gas chromatograph (SPL+wide range FID)
HS-900 semi-automatic headspace sampler (12 positions)
3097GMP version network workstation
Nitrogen: 40L steel cylinder with purity>99.999%
Air: Air Generator (XGA-2000)
Hydrogen: Hydrogen Generator (XGH-300)
Pipette gun (100ul -1000ul)
Several 50ml volumetric flasks
Ethylene oxide 10mg/ml (using water as solvent)
2-chloroethanol 10mg/ml (using water as solvent)
3. Analysis method parameters
Chromatographic parameters:
Subject Name |
Detector 1-FID |
chromatographic column |
6% cyanopropyl phenyl, 94% dimethyl polysiloxane; 30*0.53*3 |
Vaporization chamber temperature |
230℃ |
Valve box temperature |
|
Column box heating program |
40℃(2min),15℃/min to 100 ℃(1min) |
Detector type |
FID |
detector temperature |
280℃ |
gas |
40ml/min |
Assisting Gas |
400ml/min |
Compensation gas |
25ml/min |
carrier gas |
8ml/min |
Column front pressure |
|
Split ratio |
1:1 |
injection volume |
1ml |
Headspace parameters:
Temperature (℃)/Pressure |
Time (min) |
||
equilibrium temperature |
70℃ |
pressurization time |
40S |
Quantitative tube temperature |
140℃ |
Sample filling time |
20S |
Transmission tube temperature |
140℃ |
Sample balance |
|
Needle Temperature |
Needle insertion time |
||
Pressure pressure |
0.1MPa |
Cleaning Time |
180S |
Balancing Time |
40min |
||
4. Configure standard solution
a. Take 0.2ml of ethylene oxide (10mg/ml) and 0.4ml of 2-chloroethanol (10mg/ml) and add them to a 50ml volumetric flask. Add water to make up to 50ml;
The concentration is 0.2ml of ethylene oxide (40ug/ml) and 2-chloroethanol (80ug/ml) as the intermediate solution
b. Take 2.5ml of intermediate solution and 2.5ml of water, add them to a 20ml headspace bottle, seal it, and the concentration is EO: 20ug/ml ECH: 40ug/ml
c. Take 1.25l of intermediate solution and 3.75ml of water, add it to a 20ml headspace bottle and seal it. The concentration is EO: 10ug/ml ECH: 20ug/m
d. Take 0.625ml of intermediate solution and 4.375ml of water, add it to a 20ml headspace bottle and seal it. The concentration is EO: 5ug/ml ECH: 100ug/m
e. Take 0.25ml of intermediate solution and 4.75ml of water, add it to a 20ml headspace bottle and seal it. The concentration is EO: 2ug/ml ECH: 4ug/m
f. Take 0.125ml of intermediate solution and 4.875ml of water, add them to a 20ml headspace bottle and seal it. The concentration is EO: 1ug/ml ECH: 2ug/m
5. Standard sample spectrum

EO 20ug/ml ECH 40ug/ml

EO 1ug/ml ECH 2ug/ml

EO standard curve

ECH standard curve
6. Conclusion
This method uses headspace injectionChromatography, onceInjectionInFID detector simultaneously detects ethylene oxide and 2-chloroethanolThe detection limit is better than1ug/ml, Both reproducibility and linearity meet the requirements.Linear correlation coefficient of ethylene oxide analysis using this method>0.9999 or above, with a linear correlation coefficient of>0.9999 or above for the analysis of 2-chloroethanol.
IfECH residue requires a better detection limit, and it is recommended to use an ECD detector.