Shanghai Hengxi provides the following solutions for the "GB50325-2020 Indoor Environmental Pollution Control Standard for Civil Building Engineering", including 16 TVOC injection testing requirements.

1. Test substance16 types of TVOC, as shown in Figure 1-1.
Figure 1-1 16 TVOCs
2. Instrument configuration
Pre processing equipment: Shanghai Hengxi ATD-801 fully automatic secondary thermal desorption analyzer
Analytical instrument: Agilent 7820A gas chromatography (FID detector)
Sampling tube: T-C composite adsorption tube
3.1 Condition parameters of fully automatic secondary thermal desorption instrument:
3.2 GC condition parameters:
Temperature: 250 ℃; Diversion mode: diversion; Split ratio: 20:1; Carrier gas: N2.
Column specification: 50m * 0.320mm 0.25Micron; Column flow rate: 2.0ml/min.
Program heating: as shown in Figure 3-2-1.
Figure 3-2-1 Program Heating Parameters
Temperature: 250 ℃; Hydrogen flow rate: 30ml/min; Air flow rate: 300ml/min; Tail blowing: N2, 25ml/min.
4.1 16 TVOC chromatograms (low concentration 0.05mg/ml)


4.2 Linearity of Various Substances
The calibration data is as follows:

The linear chart is as follows:

Figure 4-1 n-hexane

Figure 4-2 Benzene

Figure 4-3 Trichloroethylene
Figure 4-4 Toluene
Figure 4-5 Octene

Figure 4-6 Butyl acetate
Figure 4-7 Ethylbenzene
Figure 4-8 p-xylene
Figure 4-9 o-xylene

Figure 4-10 Styrene

Figure 4-11 Nonane

Figure 4-12 Isooctanol
Figure 4-13 n-undecane
Figure 4-14 n-Tetradecane

Figure 4-15 n-Hexadecane