Operation steps for liquefied gas dimethyl ether N2000
Date: 2015-09-29Read: 1

1、 Operation steps:

  1. Turn on the hydrogen generator;
  2. Wait for the generator flow to decrease100After the following, open the liquefied gas analyzer;
  3. Open the computer, open itN2000Workstation, open the channel1Point data collection, view baseline,
  4. Wait for the temperature of the analyzer to rise40After stabilizing at ℃, the sample can be injected for analysis. When injecting, first turn the injection handle to the left, connect the sampling bag to the sample gas inlet, and squeeze the sampling bag approximately10In seconds, allow the sample gas to enter the instrument, then turn the injection handle to the right and immediately click to collect data or pressF5Stop collecting after the peak is out and see the results;
  5. When shutting down after completing the sample, first turn off the analyzer, and then turn off the hydrogen generator. When closing the computer program, first close the baseline, and then click the X button in the upper right corner to close the workstation.

II Judge Peak: 1.8The * peaks separated by left and right are air methane, and then separated by1More than a minute is propane propylene, followed by dimethyl ether, followed by isobutane, isobutene, n-butane, isobutene, and maleic acid.

Attention: The water level of the hydrogen generator should be between the upper and lower limits. If the water level is low, add distilled water in a timely manner. If one-third of the color changing silicone turns pink, replace the silicone.