▏Product Features


The GC3900C touch screen gas chromatograph is a set of highly sensitive chromatograph products that have been improved and optimized on the bus development platform. The simple and clear Chinese display interface is very easy to understand

Scholars have familiarized themselves with the product in a relatively short period of time, with clear order placement, reasonable layout, real-time display of system parameters through Flash animation, and dynamic display of graphics, making the instrument not monotonous.


▏Test Principle


Gas chromatograph is a separation and analysis tool for multi-component mixtures, which uses inert gas as the mobile phase and employs chromatographic column separation technology. When multiple analyses are conductedMaterial entryDuring the chromatographic column, due to various factors

The distribution coefficients of components in the chromatographic column are different, and the flow velocities of each component in the chromatographic column are also different. After a certain column lengthAfterwards, the mixed componentsDon'tLeaving the chromatographic column and entering the detector, after detection

Convert into electrical signals and send them to the data processing workstation to complete the analysis of the measured substanceDetection and analysis.

▏Test application

Basic Application

natural gas

Refers to all gases that naturally exist in nature,includingAtmospheric circle、Water Circlewait



▏Technical Specifications

indicator

parameter

Temperature of chromatographic column chamber

Temperature control range: room temperature ± 8 ℃ -450 ℃, with increments of 1 ℃

Temperature control accuracy: ± 0.1 ℃ - ± 0.2 ℃

Temperature control area: 10 temperature control channels

Program heating order

16th order

heating rate

0.1-40 ℃/min, increments of 0.1 ℃, initial and termination time range: 0-6000min, increments of 1min

Pressure protection method

The carrier gas has pressure protection function. When the gas pressure is below 0.1mpa, the instrument automatically cuts off the power supply, effectively protecting the color

Safe use of spectral columns

display mode

7-inch touchscreen display

Extreme temperature

Maximum protection value 450 ℃

external event

8通道外部事件

Thermal conductivity detector (TCD)

Ignition method: automatic ignition

Sensitivity:S ≥ 3500mV • ml/mg (conventional)

5000mv • ml/mg (highly sensitive) (benzene toluene solution)

Noise: ≤ 0.01mv

Baseline drift:≤0.03mV/30min

Linear range: ≥ 104

power supply

220V 50HZ 2000W

▏Product Configuration


Instrument model

GC3900C(This instrument has automatic ignition function)

quantity

1 unit

analyze samples

Reactive gas

Sample components

N2、CO2、CH4、C2H6、C3H8、i-C4H10、n-C4H10、i-C5H12、n-C5H12、C6 及C6+

configuration

Specification Model

quantity

host

GC3900PLus

room temperature± 8 ℃ -450 ℃, with increments of 1 ℃

Temperature control accuracy: ± 0.1 ℃ - ± 0.2 ℃

WenDegree gradient: ± 1%

16 stage program heating, heating rate 0.1-40 ℃/min, in increments of 0.1 ℃,

Initial and termination time range: 0-6000min, with increments of 1min.

Extreme temperature: maximum protection value 450 ℃

Injection system

Capillary injection system

2 sets

detector

Thermal conductivity detector (TCD)

Sensitivity:S ≥ 3500mV • ml/mg (conventional)

5000mv • ml/mg (highly sensitive) (benzene toluene solution)

1 piece

Baseline drift:≤0.03mV/30min

Noise: ≤ 0.01mv

workstation

Network Anti Control Version

Chinese version

1 set

chromatographic column

DC200/Chromosob

0.75m

1 piece

DC200/Chromosob

5.2m

1 piece

PN / chromosob

2.4m

1 piece

gas source

air generator

RNA-3000,3000ml/min

1 unit

hydrogen generator

RNH-300,300ml/min

1 unit

nitrogen cylinder

99.999% high-purity nitrogen with pressure reducing valve (self equipped)

1 bottle

Gas purifier

RNJ-2000, Trinity with valve switch

1 unit

Computer, printer

Computer, printer

Win7/ win10

1 set