E-mail
2757619169@qq.com
-
Phone
13969491989
-
Address
No. 98, Takeover Lane, Tengzhou City, Shandong Province
Shandong Ruide Jingke Instrument Co., Ltd
2757619169@qq.com
13969491989
No. 98, Takeover Lane, Tengzhou City, Shandong Province
Method for testing blood alcohol content
GA/T 842-2019
Release date: April 19, 2019
Implementation date: May 1, 2019
Centralized Unit: National Technical Committee for Standardization of Road Traffic Management
Established headspace gas chromatography analysis method for detecting alcohol (ethanol) content in blood
Qualitative and quantitative analysis of alcohol in personnel's blood during road traffic law enforcement activities
The national standard requires a detection limit of 1mg/100ml for ethanol in blood and a quantification limit of 5mg/100ml
Headspace analysis conditions:
1. Heating box temperature: 65 ℃
2. Quantitative ring temperature: 105 ℃
3. Transmission line temperature: 110 ℃
4. Sample bottle heating equilibrium time: 10.00min
5. Sample bottle pressurization time: 0.1min
6. Quantitative ring filling time: 0.1min
7. Quantitative ring equilibrium time: 0.05min
8. Injection time: 1.00min
Gas chromatograph analysis conditions:
Equipped with a hydrogen flame ionization detector (FID)
Chromatographic column: can refer to standard recommendations
Carrier gas: high-purity nitrogen
Detector temperature: 250 ℃
Injection port temperature: 200 ℃
Column temperature: Different chromatographic column temperatures are set differently
Suggest using a fully automatic headspace sampler in conjunction with it