Instructions for using high-speed freeze centrifuge
Date: 2020-06-23Read: 1

Instrument usage environment:

To ensure the safety of continuous use of the machine, the following factors may damage the centrifuge:

- Chemical effects

Environmental impacts, including natural ultraviolet radiation.

Corrosion and wear of protective cover components and other safety components.

1. Indoor use

2Atmospheric pressure: 860hPa-1060hPa

3. The applicable temperature range for instrument operation is+5 ℃ to+40 ℃

4. The relative temperature range suitable for instrument operation is ≤ 80%

5. Scope of application of instrument working power supplyAC230V 50HZ/60HZ

6. Adequate ventilation equipment must be installed indoors

7. There are vibrations and airflows around that do not affect performance

8. There is no conductive dust, explosive gas, or corrosive gas present in the surrounding air

Safe use of rotors:

1. Before the rotor operates, the sample should be accurately and symmetrically loaded and the test tube should be placed.

2. Centrifugal test tubes can load samples at different times, but they must be loaded symmetrically(Allowable weight error ≤ 1.5g)When loading samples asymmetrically, it is never allowed to turn on the machine.

3. When replacing the rotor, use the included wrench to loosen the screw counterclockwise, and then the rotor can be replaced.

Do not turn on the machine when the screw is not tightened on the shaft!