Revealing LOD of Particle Characterization Methods: How to Choose Accurate Detection Methods?
Date: 2024-09-26Read: 1

LOD (Limit of Detection)It refers to the minimum concentration or quantity of a substance that can be reliably detected by an analytical method, representing the sensitivity of the method.For drug delivery andNano drug delivery systemAccurately reporting LOD not only helps to determine the accuracy and repeatability of detection methods, but also relates to the detection of low concentration particles, regulatory compliance (regulatory agencies such as FDA, EMA, etc. usually require detailed reporting of LOD for particle characterization methods during drug declaration or quality control), method validation, and the credibility of scientific research results.So, can various particle characterization methods or instruments on the market provide accurate LOD? What are their detection principles and limitations? Next, let's delve deeper with the editor!

Nanoparticle tracking analysis (NTA) cannot provide traditional LOD


















Nanoparticle Tracking Analysis (NTA) It is a method of tracking and analyzing individual nanoparticlesBrownian motionA technique for measuring particle size and concentration. Although NTA can provide information on the particle size distribution and concentration of nanoparticles, it cannot provide traditional LOD due to the following reasons:

1. The signal-to-noise ratio is not constant:

NTA relies on light scattering to detect particles, and the scattering intensity of different particles varies greatly, especially when there is polydispersity in the sample. The scattering intensity of large particles often blocks the signal of small particles, resulting in larger particle size and lower concentration. Therefore, particles of different sizes may produce different signal intensities, making it difficult to define a unified LOD.

2. Influence of particle concentration:

The detection range of NTA is limited by particle concentration. When the concentration is too high, the mutual interference between particles of different sizes can lead to inaccurate analysis; When the concentration is too low, the inability to detect a sufficient number of particles can also cause sampling errors and random errors. Therefore, it is difficult to determine a low concentration detection limit that is applicable to all samples (the LOD of different samples may vary). Meanwhile, even for clear samples such as purified extracellular vesicles, their linear concentration range is usually only concentrated in one order of magnitude[1][2].

3. Limitations of Optical Systems:

The detection sensitivity of NTA also depends on the resolution and parameter settings of the optical system, especially when detecting small-sized nanoparticles. The limitations of optical resolution make it difficult to define a unified LOD for NTA technology.

Due to these factors, NTA provides a relatively quantitative result in characterizing particle size and concentration, rather than an absolute low detection limit.

揭秘颗粒表征方法的LOD:如何选择<a class=

Limitations of Dynamic Light Scattering (DLS)

Dynamic Light Scattering (DLS) Essentially, it is a method for measuring the fitted particle size of nanoparticles or macromolecules in suspension. Due to the dependence of DLS on the intensity of particle scattered light, there are usually certain limitations in determining the detection limit LOD

1. Dependence of scattering intensity:

DLS measures the average Brownian motion of all particles in the sample, therefore, the scattered signals generated by smaller particles or low concentration particles may be overwhelmed by background noise and difficult to accurately detect. Generally speaking, the LOD of DLS requires higher particle concentration and particle size range than commonly understood.

2Multi dispersibility and concentration effects:

In multi dispersed systems or samples with low concentrations, DLS is difficult to distinguish particles of different sizes, making it difficult to accurately provide a unified LOD. Specifically, the principle of DLS assumes that the samples have a monodisperse size distribution (PDI<0.1 or even smaller), which is not suitable for polydisperse samples (such as most biological samples); In addition, although the DLS principle cannot calculate concentration data, DLS measurement requires concentration, and inappropriate sample concentrations can lead to significant deviations in particle size, which are easily overlooked.

揭秘颗粒表征方法的LOD:如何选择<a class=


Overall, we can see that neither NTA nor DLS can provide a meaningful LOD. This is internationally recognizedExtracellular vesicle society(ISEV,International Society for Extracellular Vesicles)The latest guidelines also explicitly state that the LOD of the detection method needs to be disclosed, allowing others to verify the results regardless of sensitivity limitations. However, there is currently no method to obtain traceable LOD through nanoparticle tracking analysis (NTA), dynamic light scattering (DLS), or imaging flow cytometry[3].


揭秘颗粒表征方法的LOD:如何选择<a class=

Original translation:
The LOD of the detection method needs to be disclosed to allow others to verify the results regardless of sensitivity limitations. ”
The reliability of EV concentration measurement can be improved through orthogonal methods, and each method must have a definite LOD. ”
At present, there is no method to obtain traceable LOD for nanoparticle tracking analysis (NTA), dynamic light scattering (DLS), or imaging flow cytometry. ”


The Resistance Pulse Induction (RPS) method can represent a determined LOD

The Resistive Pulse Sensing (RPS) method can provide a definite LOD in characterizing the particle size and concentration of nanoparticles.RPS measures the size and quantity of particles by detecting the resistance changes caused by individual particles passing through micropores.Because it is a single particleThe detection method, RPS, can provide very low detection limits under certain conditions, especially suitable for detecting low concentration particles in diluted samples.LOD is usually associated with the size of micropores or nanopores and the level of noise in the sample.


Dependent on aperture size:

LOD is usually associated with the size of micropores and the level of noise in the sample. A smaller aperture can detect smaller particles, but it may also increase background noise, thereby affecting the accuracy of LOD.

Dependent on noise level:

The conductivity of the sample, the surface characteristics of the pores, and the stability of the background current all affect the detection limit. If the background noise is low, RPS can detect very small and low concentration particles, thereby achieving lower LOD.


Generally speaking, if background noise can be well controlled and appropriate aperture and measurement conditions can be selected in RPS testing experiments, lower LOD can be achieved.This makes RPS technology significantly advantageous in certain fields, such as detecting diluted biological particles or nanoparticles.


揭秘颗粒表征方法的LOD:如何选择<a class=




揭秘颗粒表征方法的LOD:如何选择<a class=

Ruixin Intelligent Manufacturing NanoCoulter Particle Size AnalyzerapplicationRPS technologyBreaking the limitations of optical principles, after more than ten years of micro nano processing technology precipitation,innovationPhotolithography of silicon-based solid nanopore chips extends the particle size detection range of classical Coulter technology from the micrometer level to the nanometer level. Simultaneously relying on high-sensitivity picoampere level current detection and signal noise analysis and processing, it can accurately reduce noise and analyze each through-hole particle. In the true sense of single particle detection, it can accurately analyze multi dispersed samples with wide particle size distribution, and has a wider range ofParticle size LOD (15-2000 nm) and extremely wide concentration LOD (10 μ m)6-1012 particles/mL)。



References

[1] Welsh JA, Goberdhan DCI, O'Driscoll L, et al. Minimal information for studies of extracellular vesicles (MISEV2023): From basic to advanced approaches [published correction appears in J Extracell Vesicles. 2024 May;13(5):e12451. doi: 10.1002/jev2.12451].J Extracell Vesicles. 2024; 13(2):e12404.

[2] Li Y, Zhang S, Liu C, et al. Thermophoretic glycan profiling of extracellular vesicles for triple-negative breast cancer management.Common Night.2024; 15(1):2292.

[3] Tian Y, Tian D, Peng X, Qiu H. Critical parameters to standardize the size and concentration determination of nanomaterials by nanoparticle tracking analysis.Int J Pharm. 2024; 656:124097.