Common problems and solutions of fully automatic COD analyzer
Date: 2025-12-09Read: 0
Fully automatic COD analyzerSuitable for the analysis of chemical oxygen demand in various water qualities, the product strictly follows national standards and is designed using the potassium dichromate titration principle. It automatically handles tedious digestion and titration work and can be widely used in water quality analysis in universities and research institutes, third-party testing, sewage treatment plants, water conservancy, environmental protection, petrochemical and other industries' laboratories.
Adopting a box type enclosed design with transparent observation windows, it isolates external environmental interference and prevents harmful reagents from overflowing, protecting the health and safety of operators.
The adaptive heating method dynamically adjusts the heating parameters to ensure the sensitivity and accuracy of heating temperature control.
Equipped with fault detection and alarm function, it can promptly alert in case of liquid leakage, component failure, misoperation and other faults, increasing the safety of system operation.
The common problems and solutions of the fully automatic COD analyzer are as follows:
Alarm for failure to collect reagents or water samples:
Sampling tube blockage: Check and clean or replace the blocked pipeline.
Nine way valve blockage: Disassemble, clean or replace the nine way valve.
Sampling tube leakage: Check and tighten the screws connecting the sampling tube and the nine way valve.
Unrecognized high and low position signals: Check the signal board and high and low position light sources, and replace the light sources if necessary.
No relative test reagents: Check and supplement the corresponding reagents.
Peristaltic pump tube damaged or not tightened: Check the pump tube and replace and tighten it if damaged.
Reagent reflux issue:
Nine way valve series valve: Remove, repair or replace the nine way valve.
Broken plug of digestion tank: Replace the white silk fixed black soft leather pad plug of the digestion tank.
Electromagnetic valve no response:
Poor sealing of digestion tank: Check and ensure good sealing.
Solenoid valve coil burned out: Replace the solenoid valve coil.
Poor contact of control wire: Check and replace the control wire from PLC to solenoid valve relay.