Analysis of Three Common Malfunctions in LED Illuminated Incubators
Date: 2016-03-07Read: 0
LED illuminated incubatorFully automatic control by microcomputer, touch switch, easy to operate; Programmable multi-stage control mode, temperature, humidity, light intensity, and time can be set separately during both day and night. Air duct ventilation, with soft air velocity and uniform temperature in the working room; Hollow reflective tempered laminated glass, with good insulation performance and beautiful appearance. Fully enclosed opaque lampshade, optional studio power supply, disinfection device, etc.

LED illuminated incubatorEquipped with super temperature and sensor abnormality protection function, ensuring the safety of instruments and samples; Choosing a full spectrum plant growth lamp is beneficial for plant growth and improves disease resistance. Equipped with power-off memory and automatic power-off time compensation function; Constant temperature control system, fast response, high temperature control accuracy.

Common faults of LED light cultivation box
Compressor not cooling:
1. Under balanced mode operation, the compressor is always in working condition, and the cooling effect is relatively good, and the compressor can be effectively maintained.
2. If it is a single standard operation, it will only work when the temperature drops to a certain value, and the opening and closing of doors will cause significant temperature fluctuations, which can lead to frequent starting of the compressor and shorten its service life.
Uneven temperature inside the LED illuminated incubator:
1. It may be caused by excessive concentration of emissions
2. Long term exposure to light can cause the bulb to overheat and result in uneven temperature
3. The circulating fan inside the box is not working
Ice formation at the circulating air outlet:
Every refrigeration instrument has a defrosting cycle during which the compressor does not cool. Melt the ice on the condenser by using heating elements. The melted water is likely to accumulate at the air outlet and cause freezing.