HCN-96Bacterial endotoxin constant temperature detector
Product Introduction:
Chinese Pharmacopoeia2002The annual version of the bacterial endotoxin test method requires that the reaction tube should be placed vertically37±1℃ is suitable for insulation in thermostats60±2Minutes. According to this requirement, the company has developed a special experimental equipment for bacterial endotoxin test by gel method——HCN-96Type test tube thermostat. Compared with the traditional constant temperature water bath,HCN-96The type of test tube thermostat hasThe following advantages.
Product Features:
►A simple and easy-to-use human-machine interface that displays all operational and setup information in real-time, making it easy to observe the device's operating status.
►Comes with a blue transparent cover to effectively prevent sample splashing(If using a traditional constant temperature water bath, the test tubes must be sealed, otherwise water vapor or small water droplets will enter the test tubes and cause contamination. Sealing each test tube one by one is not only troublesome and time-consuming, but also affects the accuracy of the experiment).
►The constant temperature module is divided intoATheBTheCTheDFour areas, each area can be timed independently(Each region has24Only test tube hole,Each instrument96A test tube holeThe instrument can be customized according to user needs2mlAmpoule for reaction tube and10×75mmStandard reaction test tube. The instrument has a beautiful appearance, compact size, and is convenient to use).
►The instrument adopts * heating technology and pulse temperature control technology, which makes the temperature extremely stable and uniform, and the temperature control accuracy reaches37±0.15℃, temperature difference between test tube holes ≤0.3℃。(The temperature control accuracy of the traditional constant temperature water bath is only±1The temperature distribution is uneven, with high water temperature near the heating tube and low water temperature away from the heating tube, which can easily cause differences in the reaction of horseshoe crab reagents).
►Built in software and hardware dual over temperature protection device, making it more reliable to use.
►Support automatic preheating function.
►Support automatic power outage recovery function.
►Support automatic startup function.
►The instrument comes with a temperature calibration function.
Product Parameters:
Temperature control range: |
room temperature+5°C~110°C |
Time setting: |
1min~99h59min/∞ |
temperature control accuracy: |
≤±0.3℃(@37℃) |
Display accuracy |
0.1℃ |
temperature uniformity: |
≤±0.2℃(@37℃) |
Heating time: |
≤15minute(25℃to100℃) |
Partition independent timing function: |
support |
Number of modules: |
4Block(96A hole) |
Maximum power: |
500W |
Voltage specifications: |
AC110or220V/50-60HZ |
Fuse: |
250V 3A F5×20 |
Size: |
370*254*178(mm) |
Weight: |
8KG |