Tool Preparation
Special screwdriver (flat/cross)
alcoholOr precision instrument cleaner
Dust free cloth/cotton swab
Anti-static gloves
Electronic maintenance platform (anti-static)
Safety Precautions
Keep the workbench clean and dry
Avoid environments with strong static electricity
Handle precision components with care
Possible reasons:
Low battery level
Poor battery contact
The motherboard is damp
Repair steps:
Replace with a new battery (SR44 type)
Clean the battery compartment contacts with a cotton swab dipped in alcohol
Check the motherboard for corrosion, mild corrosion can be cleaned with alcohol
Possible reasons:
The ruler has stains
Sensor is dirty
Wear of conductive strip
Repair steps:
Take out the ruler and wipe it unidirectionally with a dust-free cloth dipped in alcohol
Clean the sensor contact area (to avoid scratches)
Check the conductive strip, and gently wipe it with an eraser if there is slight wear
Possible reasons:
Oxidation of key contact points
Button film damaged
Repair steps:
Clean the button contacts with a cotton swab dipped in alcohol
If the button film is damaged, it needs to be replaced (professional accessories are required)
Possible reasons:
There are impurities in the orbit
Insufficient lubrication
Repair steps:
Clean the track with air blowing
Drip a small amount of precision instrument lubricating oil (avoid excessive)
Remove the battery
Remove the screws on the back (pay attention to the screw specifications and order)
Gently separate the shell
Record the position of each component (can be photographed)
Confirm that all components are clean
Install in sequence
Tighten the screws evenly (avoid over tightening)
Zero calibration:
Clean the measuring surface and close the caliper
Press and hold the "ZERO" button until it displays "zero"
accuracy verification:
Use standard measuring blocks
Multi point measurement verification of linearity
after useClean the measuring surface and loosen the locking screw
storePut it in a dedicated box to avoid squeezing
regularlyCheck the battery and clean the track every quarter
⚠️It is recommended to return to the factory for repair in the following situations:
LCD screen rupture
The main ruler is noticeably bent
Chip level malfunction
Depth gauge damaged
Non professionals are not allowed to disassemble sensors and display screens
Avoid using organic solvents (such as gasoline, acetone)
Basic accuracy verification is required after maintenance
Simple maintenance can solve 80% of common problems, but precise adjustments still require professional equipment. If unsure, it is recommended to contact a repair center.