According to the first part of the 2020 edition of the Chinese Pharmacopoeia, amino columns are used to analyze four types of sugars in honey. Shodex's amino column Asahipak NH2P-50 4E (specification: 4.6 mm × 250 mm, particle size: 5 µ m) can effectively separate these sugars, and the theoretical plate count calculated based on fructose is higher than the pharmacopoeia requirement of 2000.

sdxc011.jpg

Sample : 20 mg/mL each, 15 µL
1. Fructose
2. Glucose
3. Sucrose
4. Maltose

Column       : Shodex Asahipak NH2P-50 4E (4.6 mm I.D. x 250 mm)Eluent       : CH3CN/H2O=75/25Flow rate    : 1.0 mL/minDetector     : RIColumn temp. : Room temp.