GB/T 23964-2023 Triethylamine for industrial use
scope
This document specifies the requirements, test methods, inspection rules, labeling, packaging, transportation, and storage of industrial triethylamine.
This document is applicable to the industrial triethylamine produced by catalytic amination of ethanol as raw material in the presence of hydrogen gas.
Note: The molecular formula of triethylamine is C6H15N
The structural formula is
The relative molecular mass is 101.19 (based on the 2022 international relative atomic mass).
Terms and definitions
There are no terms or definitions that need to be defined in this document.
requirement
Appearance: Transparent liquid, no visible mechanical impurities.
Industrial triethylamine should meet the technical requirements in Table 1.
Test Method
Determination of moisture content
Karl Fischer Coulomb method (arbitration law)
prepare
Adjust the instrument to its working state. According to the requirements of electrolyte titration, an appropriate amount of benzoic acid or salicylic acid can be added in advance to adjust the pH of the electrolyte to 5-7. Pre titration should be performed after adding benzoic acid or salicylic acid to eliminate reagent blank.
determination
Use a microsyringe to transfer 50 μ L~100 μ L of the sample, immediately block the needle with a silicone rubber pad, weigh (accurate to 0.0001g), and follow the provisions of GB/T 6324.8.
GB/T 6324.8-2014 Test Methods for Organic Chemical Products Part 8: Determination of Moisture in Liquid Products Karl Fischer Coulombic Method
Karl Fischer direct electricity method
prepare
Adjust the instrument to its working state. According to the requirements of electrolyte titration, an appropriate amount of benzoic acid or salicylic acid can be added in advance to adjust the pH of the electrolyte to 5-7. Pre titration should be performed after adding benzoic acid or salicylic acid to eliminate reagent blank.
determination
Transfer 1.5mL~3mL of sample with a syringe, immediately block the needle with a silicone rubber pad, weigh (accurate to 0.0001g), and perform the direct electric method according to GB/T 6283.
GB/T 6283-2008 Determination of Moisture Content in Chemical Products Karl Fischer Method (General Method)
tolerance
Take the arithmetic mean of two parallel measurement results as the measurement result. The difference between the results of two parallel measurements shall not exceed 20% of their arithmetic mean.
Kyoto Electronics KEM Karl Fischer Coulombic Moisture Analyzer MKC-710S
Kyoto Electronics KEM Karl Fischer direct electric moisture meter MKV-710S