Kyoto Electronics KEM high-precision alcohol content analyzer ALM-155
Date: 2025-12-17Read: 0


High precision alcohol content tester ALM-155

GB 5009.225-2023 National Food Safety Standard Determination of Ethanol Concentration in Alcohol and Edible Alcohol
The fourth method is the U-shaped oscillating tube digital density meter method
Injecting the sample into the U-shaped oscillating tube measuring cell of the digital density meter, the oscillation frequency or period of the U-shaped tube also changes due to the change in the mass of the oscillating tube. The density of the sample at 20 ° C can be measured by using the oscillation frequency or period of the sample at 20 ° C and the instrument constant. By comparing the density of the alcohol aqueous solution with the ethanol concentration (alcohol degree), the ethanol concentration (alcohol degree) of the sample at 20 ° C can be calculated.
U-shaped oscillating tube digital density meter: analysis accuracy ± 0.00001g/mL, temperature control accuracy ± 0.02 ° C.

Working principle of high-precision alcohol content analyzer ALM-155:

The ALM-155 alcohol content measuring instrument is used to directly heat and distill alcohol samples to remove non-volatile substances from the samples. The distillate is then restored to its original volume with water, and the distillate is then drawn into the U-shaped oscillation tube of the digital alcohol concentration meter. Due to the change in sample density in the U-shaped tube, the vibration frequency will change. The instrument can automatically calculate the density of the distillate based on the vibration frequency of the sample distillate at 20 ° C. The instrument is equipped with the "Alcohol Water Solution Density and Ethanol Concentration (Alcohol Content) Comparison Table (20 ° C)" in GB 5009.225-2023, which can directly measure and display the volume percentage of ethanol concentration (Alcohol Content) in the sample.
Note: A binary mixture solution of ethanol and water can directly measure the alcohol concentration value.

The main features of the high-precision alcohol content analyzer ALM-155 are:

1. A desktop alcohol concentration meter with high accuracy, small footprint, and excellent performance.
2. Density meter: analysis accuracy ± 0.00001g/mL, temperature control accuracy ± 0.01 ° C.
3. Standard sample pump, one click start for injection and measurement, with a sample volume of only 8mL.
4. Built in Peltier temperature control, with a fixed temperature of 20 ° C. Only pure water calibration is required.
5. It can automatically store 100 sets of measurement results, and the data can be transferred to a USB flash drive or computer.
6. A table comparing the density and alcohol content of an alcohol aqueous solution, directly displaying the ethanol content.
7. Full range alcohol concentration measurement, simple operation, high accuracy, and fast measurement speed.

High precision alcohol content tester ALM-155 technical parameters:

Measurement range: Alcohol content: 0.00-100.00 vol%;
Density: 0.69937-1.24887 g/cm3;
Relative density: 0.7000-1.25000;
Brix: 0.0-50.0;
Grape alcohol content (Babo): 0.0 to 45.0;
Baume: 0.0 to 25.0.
Resolution: Alcohol content: 0.01 vol%;
Density 0.00001 g/cm3;
Relative density 0.00001;
Brix: 0.02;
Grape alcohol content (Babo): 0.02;
Baume: 0.01.
Repeatability: Alcohol content: (SD) 0.05% vol%;
Density: (SD) 0.00005 g/cm3;
Relative density: (SD) 0.00005;
Brix: (SD) 0.002;
Grape alcohol content (Babo): (SD) 0.002;
Baume: (SD) 0.001.
Measurement temperature: 20.00 ° C (fixed).
Alcohol content comparison table: Built in OIML, HMCE, AOAC, NIST, and GB 5009.225-2023 comparison tables.
Measurement time: 2-4 minutes (using standard peristaltic pump).
Sample size: approximately 8 milliliters (injection time 10 seconds).
Display: LCD liquid crystal display.
Injection method: Use a peristaltic pump or syringe for injection.
Automatic start function: Repeat frequency: 2-100.
Calibration method: Use pure water for calibration.
Computer software: SOFT-CAP (data acquisition software).
External interface: USB (USB flash drive or keyboard), RS-232C (printer and computer).
Data output: CSV format to USB drive.
Environmental conditions: temperature between 5-35 ° C, humidity below 85% RH.
Power supply: 100-240V, 50/60Hz.
Power consumption: approximately 30W.
Dimensions: 270 (width) x 402 (depth) x 163 (height) mm.
Weight: Approximately 10kg.