1. ConnectFully automatic oil contamination/oil particle size analyzerPower on, turn on the power switch, the instrument will perform a self-test. After the self-test is completed, the login interface will be displayed. Select the administrator or operator identity to log in, and then preheat for 10 minutes before testing.


2. Place the sampling bottle containing petroleum ether solution (grade below 6, or other suitable cleaning solution) on the sampling table, seal the pressure chamber, enter the maintenance interface, set the cleaning volume, and perform cleaning.


3Fully automatic oil contamination/oil particle size analyzerOperating the pressure chamber:


(1) Operation method of closed pressure chamber:


a. Hold the pressure chamber operating handle with one hand;


b. Lift the testing platform with the other hand and gently move it upwards to the stop position;


c. Rotate the operating handle clockwise to the stop position and seal the pressure chamber.


(2) Operation method for opening the pressure chamber:


a. Hold the pressure chamber operating handle with one hand and hold the detection table with the other hand;


b. Rotate the operating handle counterclockwise;


c. Open the pressure chamber and slowly lower the testing platform.


4. Yes, that's rightFully automatic oil contamination/oil particle size analyzerSet up. Set the testing frequency, testing method (automatic/manual testing), printing format, channel particle size and number of channels, system time, etc.


5. Disperse the test sample: tighten the bottle cap and place it in an ultrasonic cleaning tank (with a power of at least 4000W/m2) for at least 1 minute. The liquid in the ultrasonic cleaning tank should be consistent with the liquid level of the test sample bottle. After ultrasonic vibration, shake the sample bottle vigorously by hand for 5 minutes. Then use ultrasonic vibration again or adopt negative pressure degassing, and the time should be as short as possible. Let it stand for 2-3 seconds.


6. Enter the testing interface, select the displayed standard test type, and click the testing button to perform the test. If the viscosity of the test sample cannot be tested normally, pressure can be used to assist in the testing process.


7. After the test is completed, you can view the previous test data of the current number of groups; Can store and print test data.


8. After testing, use petroleum ether cleaning solution or other suitable solutions for cleaning operations. After cleaning, proceed to the next sample test or shut down.