CPEC310gas analyzer
calibration method

CPEC310
It is a product developed by Campbell Scientific
Closed loop vortex related system
When working in the field
Can significantly improve the measurement data
continuityandreliability
This not only depends onCPEC310Unique structural design
Also built uponCPEC310support
On the basis of automatic calibration of gas analyzer
This article mainly introduces
Calibration Method for CPEC310 Gas Analyzer
Translate into EnglishSCRUB MODULEandCO2 Standard gas cylinderorDew point generatorConnect to the trachea separatelyVALVE MODULECorresponding interfaces.
choice initial configuration--CO2 Spn Gas,inputCO2 Concentration value of standard gas.
Click on the image to view a larger image~
choiceinitial configuration--H2O Spn TDPEnter the set dew point temperature.
Set parameters and paths through CR1000KDSystem Control--Const Table:
Translate into EnglishCPEC310SCRUB, CHECK_ZRO, SET_ZRO, CHECK_CO2, SET_CO2set toTRUE;
ZRO_SPN_INTV:Calibration cycle (days)Modify according to the requirements;
ZRO_SPN_OFST:Calibration time (min)Set according to the requirements;
TIME_ZRO_SPN:Ventilation duration (sec)Modify according to the distance between the gas analyzer and the host box.
choiceRun Station—Auto Z/S on, set asTRUE.

Click on the image to view a larger image~
After the above parameter settings are completed
The system can follow the set cycle
Implement unmanned automatic online calibration
Translate into EnglishSCRUB MODULEandCO2 Standard gas cylinderorDew point generatorConnect to the trachea separatelyVALVE MODULECorresponding interfaces.
Set parameters and paths through CR1000KD
System Control--Const Table:
Translate into EnglishCPEC310SCRUB, CHECK_ZRO, SET_ZRO, CHECK_CO2, SET_CO2set toTRUE.
Save settings through CR1000KD and recompile
Program.
pathAttendant Zro/Spn--Prfrm Field Zro--Chk_Set_OptnThe value is set to“CHK_SET”.
pathAttendant Zro/Spn--Prfrm Field Zro--Pick ZRO_ALLThe value is set to“ZRO_ALL”Waiting for completion.
pathAttendant Zro/Spn--Prfrm Field CO2 Spn--Chk_Set_OptnThe value is set to“CHK_SET”.
pathAttendant Zro/Spn--Prfrm Field CO2 Spn--CO2 Spn GasThe value is set toConnect to CO2 standard gas concentration.
pathAttendant Zro/Spn--Prfrm Field CO2 Spn--Pick SPN_CO2The value is set to“SPN_CO2”Waiting for completion.
pathAttendant Zro/Spn--Prfrm Field H2O Spn--Chk_Set_OptnThe value is set to“CHK_SET”.
pathAttendant Zro/Spn--Prfrm Field H2O Spn--H2O Spn T_DPThe value is set toConnect dew point temperature.
pathAttendant Zro/Spn--Prfrm Field H2O Spn--Pick SPN_H2OThe value is set to“SPN_H2O”Waiting for completion.