Calibration Guide for CPEC310 Gas Analyzer
Date: 2025-12-16Read: 18


CPEC310gas analyzer

calibration method





11166-2000.png





CPEC310

It is a product developed by Campbell Scientific

Closed loop vortex related system

When working in the field

Can significantly improve the measurement data

continuityandreliability

This not only depends onCPEC310Unique structural design

Also built uponCPEC310support

On the basis of automatic calibration of gas analyzer

This article mainly introduces

Calibration Method for CPEC310 Gas Analyzer





System startupBefore the unmanned automatic calibration function

It is necessary to set up according to the following steps





  1. Translate into EnglishSCRUB MODULEandCO2 Standard gas cylinderorDew point generatorConnect to the trachea separatelyVALVE MODULECorresponding interfaces.




  1. choice initial configuration--CO2 Spn GasinputCO2 Concentration value of standard gas.


WXWorkCapture_1765350397344.png

WXWorkCapture_176535040185.png

Click on the image to view a larger image~




  1. choiceinitial configuration--H2O Spn TDPEnter the set dew point temperature.




  1. Set parameters and paths through CR1000KDSystem Control--Const Table

  1. Translate into EnglishCPEC310SCRUB, CHECK_ZRO, SET_ZRO, CHECK_CO2, SET_CO2set toTRUE

  2. ZRO_SPN_INTV:Calibration cycle (days)Modify according to the requirements;

  3. ZRO_SPN_OFST:Calibration time (min)Set according to the requirements;

  4. TIME_ZRO_SPN:Ventilation duration (sec)Modify according to the distance between the gas analyzer and the host box.




  1. choiceRun Station—Auto Z/S on, set asTRUE.

WXWorkCapture_17653503908942.png

Click on the image to view a larger image~





After the above parameter settings are completed

The system can follow the set cycle

Implement unmanned automatic online calibration







Of course!

The system also supports calibration through manual methods




  1. Translate into EnglishSCRUB MODULEandCO2 Standard gas cylinderorDew point generatorConnect to the trachea separatelyVALVE MODULECorresponding interfaces.

  1. Set parameters and paths through CR1000KD

    System Control--Const Table

  1. Translate into EnglishCPEC310SCRUB, CHECK_ZRO, SET_ZRO, CHECK_CO2, SET_CO2set toTRUE.

  2. Save settings through CR1000KD and recompile

    Program.

  1. pathAttendant Zro/Spn--Prfrm Field Zro--Chk_Set_OptnThe value is set to“CHK_SET”.

  1. pathAttendant Zro/Spn--Prfrm Field Zro--Pick ZRO_ALLThe value is set to“ZRO_ALL”Waiting for completion.

  1. pathAttendant Zro/Spn--Prfrm Field CO2 Spn--Chk_Set_OptnThe value is set to“CHK_SET”.

  1. pathAttendant Zro/Spn--Prfrm Field CO2 Spn--CO2 Spn GasThe value is set toConnect to CO2 standard gas concentration.

  1. pathAttendant Zro/Spn--Prfrm Field CO2 Spn--Pick SPN_CO2The value is set to“SPN_CO2Waiting for completion.

  1. pathAttendant Zro/Spn--Prfrm Field H2O Spn--Chk_Set_OptnThe value is set to“CHK_SET”.

  1. pathAttendant Zro/Spn--Prfrm Field H2O Spn--H2O Spn T_DPThe value is set toConnect dew point temperature.

  1. pathAttendant Zro/Spn--Prfrm Field H2O Spn--Pick SPN_H2OThe value is set to“SPN_H2O”Waiting for completion.